Input Ratings

Power type:  Single-phase ; Three-phase, Delta or Y connection selectable

Voltage range:  200 Vac to 240 Vac ±10 % phase voltage (Delta: L-L, Y: L-N)

Frequency range: 47 Hz to 63 Hz

Power factor: 0.95 or higher (typ.)

Efficiency: 80 % or higher

Maximum power consumption: 6 kVA or lower


AC Output: Multi-phase output

*Single-phase output 

Output capacity: 4.5kVA

Mode: IP2W

Setting mode: --

Phase voltage:

 - Setting range:0.00 V to 175.0 V / 0.0 V to 350.0 V (sine and square wave), Setting Resolution: 0.01 V / 0.1 V

                 0.00 Vpp to 500.0 Vpp / 0.00 Vpp to 1000 Vpp (triangle and arbitrary wave), Setting Resolution: 0.01 Vpp / 0.1 Vpp / 1 Vpp

 - Accuracy: ±(0.3 % of set + 0.5 V / 1 V)

Line Voltage Setting range: --

Maximum current: 45A/ 22.5A

Maximum peak current: Four times of the maximum RMS current

Load power factor: 0 to 1 (leading phase or lagging phase, 45 Hz to 65Hz)

Frequency:

  - Setting range: AC Mode: 15.00 Hz to 2000.0 Hz, AC+DC Mode: 1.00 Hz to 2000.0 Hz, Setting resolution: 0.01 Hz / 0.1 Hz

  - Accuracy: ± 0.01% of set

  - Stability: ± 0.005%

Output on phase setting range: 0.0° to 359.9° variable (Free / Fix selectable), 0.1° (1 Hz to 500 Hz), 1° (500 Hz to 2000 Hz)

Output off phase setting range: 0.0° to 359.9° variable (Free / Fix selectable), 0.1° (1 Hz to 500 Hz), 1° (500 Hz to 2000 Hz)

Setting range of the phase angel: --

Phase angle accuracy: --

DC offset:± 20 mV (typ.)

* Polyphase output 

Output capacity: 1P3W: 3 kVA ; 3P4W: 4.5 kVA

Mode: 1P3W ; 3P4W (Y-connection)

Setting mode:  Independ, Balanced

Phase voltage:

 - Setting range:0.00 V to 175.0 V / 0.0 V to 350.0 V (sine and square wave), Setting Resolution: 0.01 V / 0.1 V

                         0.00 Vpp to 500.0 Vpp / 0.00 Vpp to 1000 Vpp (triangle and arbitrary wave), Setting Resolution: 0.01 Vpp / 0.1 Vpp / 1 Vpp

 - Accuracy: ±(0.3 % of set + 0.5 V / 1 V)

Line Voltage Setting range: 

   - 1P3W: 0.00 V to 350.0 V /0.00 V to 700.0 V3P4W: 0.00 V to 303.1 V /0.00 V to 606.2 V(sine and square wave)

      Setting Resolution: 0.01 V /0.1 V

   - 1P3W: 0.00 Vpp to 1000 Vpp /0.00 Vpp to 2000 Vpp3P4W: 0.00 Vpp to 866.0 Vpp /0.00 Vpp to 1732 Vpp(triangle and arbitrary wave)

      Setting Resolution: 0.01 Vpp /0.1 Vpp / 1 Vpp

Maximum current: 15 A / 7.5 A

Maximum peak current: Four times of the maximum RMS current

Load power factor: 0 to 1 (leading phase or lagging phase, 45 Hz to 65Hz)

Frequency:

  - Setting range: AC Mode: 15.00 Hz to 2000.0 Hz, AC+DC Mode: 1.00 Hz to 2000.0 Hz, Setting resolution: 0.01 Hz / 0.1 Hz

  - Accuracy: ± 0.01% of set

  - Stability: ± 0.005%

Output on phase setting range: 0.0° to 359.9° variable (Free / Fix selectable), 0.1° (1 Hz to 500 Hz), 1° (500 Hz to 2000 Hz)

Output off phase setting range: 0.0° to 359.9° variable (Free / Fix selectable), 0.1° (1 Hz to 500 Hz), 1° (500 Hz to 2000 Hz)

Setting range of the phase angel: (polyphase output)

   - 1P3W: L2 phase: 0° to 359.9° 

   - 3P4W: L2 phase: 0° to 359.9°

                 L3 phase: 0° to 359.9°

     Setting Resolution: 0.1°

Phase angle accuracy: (polyphase output)

45 Hz to 65 Hz: ±1.0°

15 Hz to 2000 Hz: ±2.0° 

DC offset:± 20 mV (typ.)


DC Output ( Only single Phase output)

Output capacity: 4.5kW

Mode: Floating output, the N terminal can be grounded

Voltage:

  -Setting range:-250.0 V to +250.0 V / -500.0 V to +500.0 V, Setting Resolution: 0.01 V / 0.1 V

  - Accuracy: ±(|0.3 % of set| + 0.3 V / 0.6 V)

Maximum Current: 45 A / 22.5 A

Maximum peak current: Four times of the maximum current


Output stability, total Harmonic Disortion, Output voltage time and Ripple noise

Line regulation: ±0.1% or less (Phase voltage)

Load regulation:

±0.1 V / ±0.2 V, @DC (only single-phase output)

±0.1 V / ±0.2 V, @45 Hz to 65 Hz (phase voltage, 0 to 100%, via output terminal)

±0.5 V / ±1.0 V, @all other frequencies (phase voltage, 0 to 100%, via output terminal)

Disortion of output: <0.3 % @1Hz to 100Hz, <0.5 % @100.1 Hz to 500 Hz, <1 % @500.1 Hz to 2000 Hz

Output  voltage respone time: Fast: 50 µs (typ.) ; Middle: 100 µs (typ.) ; Slow: 300 µs (typ.)

Ripple noise: 0.5 Vrms / 1 Vrms (TYP)



Standard Accessories

QuickStart Guide

Safety guide

Input terminal cover

Output terminal cover

Copper plate for delta connection input (Mark 1)

Copper plate for single phase and Y connection input (Mark 2)

Copper plate for delta connection input (Mark 3)

Copper plate for 1P output (Mark 4)

GRA-451-E Rack mount adapter (EIA)

USB cable (USB 2.0 Type A - Type B cable, approx. 1.2M)

Details