For full functionality of this site it is necessary to enable JavaScript.

Advanced Energy Mikron M345X6-LC Calibration Source (-40 ~ 100°C)

- Temperature Range: -40 ~ 100°C (-40 ~ 212°F)
- Temperature Uncertainty: ±0.05°C (thermometric calibration)
- Temperature Resolution: 0.01°C
- Stability: 0.02°C per 8-hour period with no external air breezes
- Source Non-Uniformity: ±0.2°C (excluding 10% border area) up to 50°C
- Heated Cavity Shape: Thermally uniform plate
- Exit Port Diameter: 152 x 152 mm (6 x 6 in)
- Emissivity ε:
0.9756 ±0.0039 at 8 ~ 15 μm
0.9713 ±0.0049 at 3 ~ 5 μm
- Calibration Method:
Thermometric
- Temperature Sensor: Precision platinum RTD 1/3 DIN
- Warm-up Time: 15 minutes (-40°C)
- Slew Rate to 1°C Stability: 15 minutes max to -40°C with chiller set at 0°C
- Slew Rate to 0.1°C Stability: 15 minutes max for entire range travel
Environmental
- Operating Ambient Temp: 20 ~ 30°C
- Cooling: Fan cooled, air inlet/outlet on top panel
- Operating Humidity: Dew point must be lower than blackbody setpoint
- Dimension (H x W x D):
Controller module: 178 x 483 x 593 mm (7 x 19 x 23 in)
- CE Certified: Yes
Communication and Electrical Specifications
- Remote Set Point: RS232
- Method of Control: Digital PID controller
- Power Requirements: 230 VAC ±10% 50 Hz
Accessory
Option: Serial Communication Output RS485

Details

Datasheet

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information