Ainuo AN9651TH (F) Three Phase Electrical Safety Comprehensive tester

1) Ground Bond Test
Output Current Setting Range: AC3A~30A 50Hz/60Hz,1A/step
Upper Limit of Resistance: 3A≤output current≤10A, 10~600mΩ; 10A
Ground Bond Measurement Range: 10~600mΩ                                                                                                                                                                                                                                                                  
Compensation Resistance Measurement Range: 0~200mΩ 1mΩ
Test Time Setting Range: 0.5~999.9s 0.1s/step
2) Insulation Resistance Test
Rated Output Load/Current: 1VA (1000V/1mA) 1mA
Output Voltage Range: DC:500V/1000V <1%
Insulation Resistance Range: 1.0~2000MΩ
Insulation Resistance Lower Limit Setting Range: 1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step
3) AC Hipot Test
Rated Output Load: 500VA (AC:5000V/100mA)
Output Voltage Range /Frequency/Adjustable Rate: 500~5000V, 50Hz/60Hz, <2%
Leakage Current Preset Alarm: Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step
Leakage Current Measurement Range: 0.10~99.99 mA
4) DC Hipot Test
Rated Output Load: 60VA (AC:6000V/10mA)
Output Voltage Range /Frequency/Adjustable Rate: 500~6000V
Leakage Current Preset Alarm: Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA    
Leakage Current Measurement Range: 50~9999μA 
5) Leakage Current Test
Leakage Current Range/Resolution: Upper limit: 50~9999μA 1μA
Compensation Current Setting Range/ Resolution: 0~500μA 1μA
Leakage Type: dynamic, static
Three Phase Dynamic Mode: A mode: ABC-AB-BC-CA
6) Power Test
Power Setting Range/Resolution
+ Upper limit, Single phase:20~6000W Three phase:60~20000W
+ Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step
Voltage Measurement Range and Accuracy: AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
Current Measurement Range and Accuracy: AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)
Active Power Measurement Range and Accuracy: 20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW; 45~65Hz: ±(0.5% of reading +2counts) 
7) Low Voltage Start Test
Upper limit: 0.30 ~30.00A 
Lower limit: 0.00 ~30.00A0.01A
Voltage Measurement Range and Accuracy: AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)
Current Measurement Range and Accuracy: AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts) 
 Isolated Transformer Capacity: Three Phase:20kVA, Single Phase: 6kVA 
 DUT Type Setting: Single Phase, Three phase Three line, Three phase Four line 
Dimension (WxHxD): 765x1766x665
Gross weight: 260kg

Details

 

Product Detail

Sigle-phase/Up to 7 functions
Double load for paralleled testing; 7 functions in 1 electrical safety and properties’ test; double channel display together; electric ARC, load open-circuit and short-circuit test function.
Applicable to such small home appliances high-speed product lines.

Technical characteristics:
Compliant with GB4706.1-2005, IEC60335-1:2004, UL60335-1:2002, GB4943.1-2001, standard requirements.
Three phasesingle phase load compatible; Three phase rated load capacity 20kVA, can custom 60kVA.
Seven functions into one device: ground bond, insulation resistance, AC hipot, DC hipot, leakage current, power test and low voltage start test.
15-group / 8-step test condition, intelligent judgment, auto compensation, auto protection;High speed test, the fastest speed can be 0.5s.
Voltage ramp-up and ramp-down, discharge function;Dynamic and static leakage current, hipot test.
Auto compensation, auto protection, auto memory, auto judgment, voice and light alarm.
Serial communication, print, foot pedal interface, outer alarm light, PLC, network card interface.
For safety test, the precision is 2%, for power test, the precision is 5%;For DUT, can support both grounding and suspending.
Function preview, single screen setting, no need flip, simple operation.

Support both 50Hz and 60Hz frequency, Chinese and English operation interface.

Model

AN9651TH (F)(Three Phase/single phase7 in1)

Ground Bond Test

Empty Load Output Voltage

<12v span="">

Output Current Setting Range

AC3A~30A 50Hz/60Hz,1A/step

Output Current Setting Accuracy

±(2% of setting+0.2A)

Upper Limit of Resistance

3A≤output current≤10A, 10~600mΩ; 10A

Ground Bond Resolution/ Measurement Range/ Accuracy

1mΩ, 10~600mΩ, ±(2% of reading +2mΩ)

Compensation Resistance Measurement Range/Accuracy

0~200mΩ 1mΩ

Test Time Setting Range

0.5~999.9s 0.1s/step

Insulation Resistance Test

Rated Output Load/Current

1VA (1000V/1mA) 1mA

Output Voltage Range/Adjustable

DC:500V/1000V <1% (Empty load to full load)

Output Voltage Ripple

Pure resistance load, 1000V/1mA, <3%

Insulation Resistance Range

1.0~2000MΩ

Insulation Resistance Lower Limit Setting Range/ Resolution

1.0~99.9 MΩ, 0.1MΩ/step, 100~2000 MΩ 1 MΩ/step

Insulation Resistance Accuracy

1~200 MΩ: ±(2% of reading+2counts) , 200~2000 MΩ: ±(5% of reading+5counts)

Ramp-up Time Range/Resolution

0.1~999.9s 0.1s

Delay Judgment Time

0.3~10.0s (Count from ramp-up )

Test Time Setting Range

0.5~999.9s 0.1s/step

AC Hipot Test

Rated Output Load

500VA (AC:5000V/100mA)

Output Voltage Range /Frequency/Adjustable Rate

500~5000V, 50Hz/60Hz, <2%(Empty load to full load)

Output Voltage Setting Range/Accuracy

Cold state AC:500~5000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)

Leakage Current Preset Alarm

Upper limit: 0.10~99.99mA; Lower limit: 0.00~99.99mA , 0.01mA/step

Compensation Current Setting Range/ Resolution

0.00 ~5.00mA, 0.01mA

Output Initial Voltage Range

(0%~99%) of test voltage, from 0V to initial voltage need 200ms

Leakage Current Measurement Range/Resolution/Accuracy

0.10~99.99 mA, 0.01mA, ±(2% of reading +2counts)

Test Time/Ramp-Up/Ramp-Down

0.5~999.9s, 0.1~999.9s, 0.0~999.9s(After test pass)

DC Hipot Test

Rated Output Load

60VA (AC:6000V/10mA)

Output Voltage Range /Frequency/Adjustable Rate

500~6000V, <2%(Empty load to full load)

Output Voltage Setting Range/Accuracy

Cold state AC:500~6000V, Hot state AC:500~2500V, 5V/step, ±(2.5% of setting+5V)

Leakage Current Preset Alarm

Upper limit: 50~9999μA; Lower limit: 0~9999μA , 1μA

Compensation Current Setting Range/ Resolution

0 ~5000μA, 1μA

Output Initial Voltage Range

(0~99%) of test voltage, from 0V to initial voltage need 200ms

Leakage Current Measurement Range/Resolution/Accuracy

50~9999μA , 1μA, ±(2% of reading +2counts)

Test Time/Ramp-Up/Ramp-Down

0.5~999.9s, 0.1~999.9s, 0.3~10.0s(Count from ramp-up)

Leakage Current Test

Output Voltage Range

1.06 times of input voltage (output by isolated transformer)

Leakage Current Range/Resolution

Upper limit: 50~9999μA 1μA

Compensation Current Setting Range/ Resolution

0~500μA 1μA

Test Time Range/Resolution

1.0 ~999.9s 0.1s

Leakage Type

dynamic, static

Three Phase Dynamic Mode

A mode: ABC-AB-BC-CA (standard)
B mode: ABC-AB-ABC-BC-ABC-CA (customized)

Output Voltage Measurement Range/Accuracy

AC 50~500.0V 0.1V, ±(0.5% of reading+2counts)

Leakage Current Measurement Range/Accuracy

50~9999μA 1μA
DC~10kHz: ±(2% of reading+5counts) 10kHz~1MHz: ±(5% of reading+5counts)

Leakage Current Accuracy

50.0~2000μA is ±(5% of reading);2000μA or above is ±(2.5% of reading)

Leakage Power

6kVA ,isolated transformer capacity is designed according to the rated voltage 220V

Time Setting Range

1~300s

Measurement Mode

dynamic, static

Power Test

Output Voltage Range

1.00 times of input voltage (output by isolated transformer)

Power Setting Range/Resolution

Upper limit, Single phase:20~6000W Three phase:60~20000W
Lower limit, Single phase:0~6000W Three phase:0~20000W 1W/step

Test Time Range/Resolution

0.5 ~999.9s 0.1s

Voltage Measurement Range and Accuracy

AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)

Current Measurement Range and Accuracy

AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)

Active Power Measurement Range and Accuracy

20.0~999.9W; 1.000~9.999kW; 10.00~20.00kW
45~65Hz: ±(0.5% of reading +2counts)

Power Factor Range and Accuracy

0.100~1.000, ±(2% of reading +2counts)

Low Voltage Start Test

Output Voltage Range

0.85 times of input voltage (output by isolated transformer)

Test Time Range/Resolution

0.5 ~999.9s 0.1s

Current Setting Range/Resolution

Upper limit: 0.30 ~30.00A Lower limit: 0.00 ~30.00A0.01A

Voltage Measurement Range and Accuracy

AC:50.0V~500.0V, 45~65Hz: ±(0.5% of reading +2counts)

Current Measurement Range and Accuracy

AC:0.30~30.00A, 45~65Hz: ±(0.5% of reading +2counts)

Display Mode

10.4 Inch, 640×480, TFT LCD display

Isolated Transformer Capacity

Three Phase:20kVA, Single Phase: 6kVA

Over Current Protection

Power, Start, Leakage Current, Phase Current35A, over 5s protection

DUT Type Setting

Single Phase, Three phase Three line, Three phase Four line

<span style="font-size:9.0pt; font-f