Fluke 1502A Thermometer Readouts (−200 °C to 962 °C, ±0.012 °C)
Manufacturer: Fluke (Calibration) Model: 1502A - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best priceHanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222)730.39.68
HCM city: (028) 38.119.636
Temperature Range: −200 °C to 962 °C (−328 °F to 1764 °F)
Resistance Range: 0 Ω to 400 Ω, auto-ranging
Probe: Nominal RTPW: 10 Ω to 100 Ω, RTD, PRT, or SPRT
Resistance Accuracy:
0 Ω to 20 Ω: 0.0005 Ω
20 Ω to 400 Ω: 25 ppm
Temperature Accuracy:
±0.004 °C at −100 °C
±0.006 °C at 0 °C
±0.009 °C at 100 °C
±0.012 °C at 200 °C
±0.018 °C at 400 °C
±0.024 °C at 600 °C
Resistance Resolution:
0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω
Temperature Resolution: 0.001 °C
Probe Connection: 4-wire with shield, 5-pin DIN connector
Communications: RS-232 serial standard
Display: 8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters
Power: 230 VAC (±10 %), 50/60 Hz, 1 A, nominal, specify
Size (HxWxD): 61 x 143 x 181mm (2.4 x 5.6 x 7.1 in)
Probes from Fluke Calibration: 5615, 5627, 5626, 5628, 5622 (option)
- Quality Engagement
- Easy change and return
- Delivery Avaliable
- Favorable payment