HV HIPOT GD3126E Insulation Resistance Tester (15kV)

Rated test voltage steps: 0.5kV, 1.0kV, 2.0kV, 2.5kV, 5.0kV, 10kV, 15kV

Output voltage error: ±5%

Resistance range

- Upper limit: 100GΩ~5TΩ

- Lower limit: 0.1MΩ

Accuracy class: 5.0 class

Lower half range basic error: ±( 5%•Rx + 1d )

Upper half range basic error: ±( 10%•Rx + 2d )

High voltage display error: ±( 5%•Ux + 1d )

Temperature measuring error: ±0.5℃

Humidity measuring error: ±2%RH

Short circuit current: Approx. 5mA

Voltage ripple: <3%

Insulation resistance: > 500 MΩ (between test wire and case)

Withstand voltage: AC10.0kV 50Hz 1min (between test wire and case)

Power supply: 14.8V lithium battery

Battery service time after one charging: 30days 10times DAR & 5times PI test per day

Working temperature and humidity: 0℃ ~ + 40℃   < 85%RH

Storage temperature and humidity: - 20℃ ~ + 50℃  < 90%RH

Dimension (mm): 460 (L)*380(W)*190(H)

Weight: 8.5kg