For full functionality of this site it is necessary to enable JavaScript.

Memmert HCP108 Humidity chamber

Temperature
Working temperature range     with humidity min. 8°C above ambient up to +90°C
Working temperature range     without humidity min. 8°C above ambient up to +160°C
Temperature     2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Display     resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED)
Humidity
Humidity     active humidifying and de-humidifying control (20-95 %) with digital display of relative humidity - resolution of display 0.5 %, setting accuracy 1 %
Humidity     humidity supply with distilled water from external tank by self-priming pump
Humidification     humidification by hot steam generator
Control technology
Display     digital display of all set parameters, such a temperature, weekdays, time, CO2, humidity and set-up values - language to be chosen via set-up
Controller     Electronic microprocessor temperature controller with auto-diagnostic system
Timer     integrated timer for tempering profiles of up to 40 ramps each, each segment adjustable from 1 min. to 999 hrs.
Calibration     three freely selectable temperature values
Communication
Interface USB     USB-interface incl. Memmert software "Celsius" for programming and documentation
Interface Printer     parallel printer interface (incl. real time clock with date function) for all PCL3-compatible ink jet printers for GLP-conforming documentation
Documentation     integrated ring memory as data logger for GLP-conforming long-term documentation of all relevant parameters - 1024 kB
Documentation     programme stored in case of power failure
Programming     chip-card control incl. 1 MEMoryCard XL with 32 kB storage capacity (max. 40 ramps)
Safety
Temperature control     mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control     overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY     additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system     integral fault diagnostics for temperature and humidity control
Alarm     with visual and acoustic alarm in case of over-/under temperature and underhumidity, open door and empty water tank
Heating concept
6 sides     large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Ventilation     uniform atmosphere and temperature distribution owing to enclosed non-turbulent ventilation system in working chamber