Memmert IF55mplus Incubator Im

-  Volume: 53 liters

-  Interrior dimensions: 400W x 400H x 330D mm

-  Exterior dimensions: 585W x 784H x 514D mm

-  Stainless steel grids: 1

-  Max. number of grids: 4

-  Max. loading per grid: 20 kg

-  Max. loading of chamber: 80 kg

-  Working temperature range: min. 10 above ambient temperature to 80°C

-  Setting accuracy: 0.1°C

-  Forced convection

-  Two high-resolution TFT colour displays with touch-sensitive buttons for selection of functions.

-  Available parameters on the ControlCOCKPIT: temperature (Celsius or Fahrenheit), exhaust air flap position, programme time

-  Two Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error

-  AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

-  ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function

-  Displaying of already logged protocol data on the ControlCOCKPIT (max 10000 values correspond to approx. 1 week)

-  Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading programmes and for online logging

-  Multiple overtemperature protection: Electronic temperature monitoring TWW/TWB (protection class 3.1) and mechanical temperature limiter TB (protection class 1) acc. to DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over/undertemperature alarm and also for all other parameters

-  PID microprocessor control with integrated auto-diagnostic system

-  Structured stainless steel housing, scratch-resistant, robust and durable; rear of zinc-plated steel

-  Internal data logger with a storage capacity of at least 10 years

-  Language setting: German, English, Spanish, French, Polish, Czech, Hungarian

-  Timer: digital backwards counter with target time setting, adjustable from 1 minute to 99 days

-  Function setpointWAIT: the process time does not start until the set temperature is reached

-  Calibration: three freely selectable values each

-  Programme stored in case of power failure

-  Autodiagnostic system for fault analysis

-  Door: fully insulated stainless steel door with 2-point locking, inner glass door

-  Ambient temperature: 5 – 40°C

-  Humidity: max. 80% rh, non-condensing

-  Voltage: 230 V, 50/60 Hz

- Electrical load: 1000 W


Accessories:

+ Works calibration certificate at 37°C (Memmert supply)

+ Stainless steel grid, 1 pc

+ Manual