PRESYS MCS-12 Universal Calibrator

Input Ranges

Resistance:

0 to 400 Ω; Resolution: 0.01 Ω; Accuracy: ± 0.01 % FS; Remarks: Excitation current 0.85 mA,auto-ranging

400 to 2500 Ω; Resolution: 0.01 Ω; Accuracy: ± 0.03 % FS; Remarks: Excitation current 0.85 mA,auto-ranging

Frequency:

0 to 600 Ω; Resolution: 0.01 Ω; Accuracy: ± 0.02 % FS; Remarks: R > 50 kΩ

600 to 1300 Hz; Resolution: 0.1 Hz; Accuracy: ± 0.2 Hz; Remarks: Voltage DC = 30 V

1300 to 10000 Hz; Resolution: 1 Hz; Accuracy: ± 2 Hz; Remarks:AC Signal from 0.3 to 30 V auto-ranging

Outputs Ranges

Resistance:

0 to 400 Ω; Resolution: 0.01 Ω; Accuracy: ± 0.02 % FS; Remarks: For external excitation current of 1.0 mA

0 to 2500 Ω; Resolution: 0.1 Ω; Accuracy: ± 0.03 % FS; Remarks: For external excitation current of 1.0 mA

Frequency:

0 to 100 Hz; Resolution: 0.01 Hz; Accuracy: ± 0.02 Hz; Remarks:Peak value:22 V / 25 mA max.

0 to 10000 Hz; Resolution: 1 Hz; Accuracy: ± 2 Hz; Remarks:Peak value:22 V / 25 mA max.


Details

Datasheet 

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information