TESTO NanoMet3 Portable Nanoparticle Counter for Real Driving Emissions (RDE)
Manufacturer: TESTO Model: NanoMet3 P/N: 444 Origin: Germany Guarantee: 12 Month - Request a Quotation
- Contact
See more: Equipment Calibration-Inspection & Repair Service
Call for the best priceHanoi city: (024) 35.381.269
Danang city: (023) 63.747.711
Bac Ninh city: (0222)730.39.68
HCM city: (028) 38.119.636
Aerosol: Exhaust gas or air with nanoparticles
Concentration range Sensor: 1E3 bis 1E6 pt/cm3; diluted: 1E4 bis 3E8 pt/cm Particle size: 10 bis 700 nm = 0.01 to 0.70 μm
Range of average particle size (modal diameter): 10 bis 300 nm = 0.01 to 0.30 μm
Raw gas flow: 4.0 lN/min, actively transported to diluter by internal pump
Dilution factors Standard: 10, 100, 300
(freely selectable or automatic selection by the measurement system)
Measurement gas flow: 1.0 lN/min
Power supply: 12 to 24 VDC, max. 60 A. 90 – 240 VA 50/60 Hz
Power consumption: Nominal power 650 W; 300 W under standard ambient conditions
Evaporator pipe temperatures: From ambient temperature up to 300 °C; accuracy ±3 °C
Assembly: 19” housing with carrying handles
Weight: Approx. 18 kg; with accessories connected approx. 23 kg
Dimensions (W x H x D): 526 x 409 x 216 mm
Operating conditions:
Ambient temperature: 5 to 35 °C; 0 to 80 % relative humidity, max. 80 %
at 30 °C, linear reduction down to 50 % at 35 °C, non-condensing
Sensor calibration: Standard calibration with NaCl particles
System calibration: Against PMP system with soot from particle generator with GMD 60 nm and 85 nm