Main Unit Specifications

Item

Description

Number of analog input channels

10 channels

External Input/Output

Trigger input, Logic input 4 ch or Pulse input 4 ch, Alarm output 4 ch

Sampling interval

100 ms to 1 h (Max:100 ms/10ch)

TIME/DIV

1 sec/DIV to 72 h/DIV

Trigger function

Type

Start

Data capture starts when a trigger is generated

Stop

Data capture starts when a trigger is generated

Condition

Start

Off, Level, Alarm, Scheduled time, External, Day

Stop

Off, Level, Alarm, Scheduled time, Time, Day, External

Condition

Analog

H, L, Window In, Window Out, Rate of change

Logic

H, L

Pulse

H, L, Window In, Window Out, Rate of change

Alarm output

No. of channels

4

Output format

Relay contact output (NO/NC)

Output conditions

Level judgement, Window judgement, Logic pattern judgement, Pulse judgement

Output terminal

M4 screw terminal

Rating

250 VAC/2A, 30 VDC/2A

Pulse input range

Count mode

50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S. (max. 50 k/sampling interval)

Inst. Mode

50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S. (max. 50 k/sampling interval)

RPM mode

50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M RPM/F.S. (max. 50 k/sec)

Calculation function

Statistics calculation *1

Average, Peak, maximum, Minimum, RMS (2 calculations can be set simultaneously)

Calculation between channels

Addition, Subtraction, Multiplication, Division (for analog input ch1 to ch10)

Wet-and dry-bulb temperature conversion

Humidity is calculated from the measured temperature of Wet-bulb and Dry-bulb
(Dry bulb : fixed to ch1, Wet bulb : fixed to ch2, Range 0 to 100 %)

Other function

Searching, Annotation input, Message, Marker function

Interface to PC

Ethernet (10BASE-T / 100BASE-TX), USB (Compatible with full speed)

Ethernet function

Web server function, FTP server function, FTP client function, NTP client function

USB function

USB drive mode (File transfer/delete of internal memory)

Internal storage device

Internal flash memory : 14 Mbyte, USB memory slot (compatible with full speed) *2

Display

Size

5.7 inch TFT color LCD

Displayed items

Waveforms + Digital values (vertical, horizontal), Digital + Calculation, Bar graph (vertical)

Clock accuracy (23°C environment)

±0.002 % (within approx. 50 sec per month)

Operating environment

Temperature : 0 to 50 °C, Humidity : 5 to 85 % R.H.

Withstand voltage

Between each input channel and GND: 1 minute at 350 Vp-p

Between input terminals: 1 minute at 350 Vp-p

Power supply

100 to 240 V AC / 50 to 60 Hz, M4 screw terminal (Power cord is not attached)

Power consumption

38 VA (when LCD is ON)

External dimensions (W x H x D)(approx.)

144 mm x 144 mm x 200 mm

Weight (approx.)

2.1 kg

Vibration resistance 

Compatible with JIS Vibration testing methods for automobile Type 1 Class A-equivalent

Dust proof, Drip proof: IP65 compatible (for front panel only)

Analog input Specifications

<td style="border:solid windowtext 1.0pt;mso-border-alt:solid windowtext .5pt;

Item

Description

Type of input terminal

M4 screw type terminal

Measurement range

Voltage

20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, V, 1-5 V F.S.

Temperature

Thermocouple : K, J, E, T, R, S, B, N, W (WRe5-26)

R.T.D. : Pt100, JPt100, Pt1000 (IEC751)

Humidity

0 to 100 % RH (Voltage 0 to 1 V conversion, when using optional B-530 humidity sensor is used)

Input filter

Off, 2, 5, 10, 20, 40 (Moving average)

Measurement accuracy *3

Voltage

±0.1 % of F.S.

Thermocouple

Measurement range

Accuracy

R/S

0 TS 100 °C
100 < TS 300 °C
R : 300 < TS 1600 °C
S : 300 < TS 1760 °C

± 5.2 °C
± 3.0 °C
± (0.05 % of rdg + 2.0 °C)
± (0.05 % of rdg + 2.0 °C)

B

400 TS 600 °C
600 < TS 1820 °C

± 3.5 °C
± (0.05 % of rdg + 2.0 °C)

K

-200 TS -100 °C
-100 < TS 1370 °C

 

E

-200 TS -100 °C
-100 < TS 800 °C

 

T

-200 TS -100 °C
-100 < TS 400 °C

 

J

-200 TS -100 °C
-100 < TS 100 °C
100 < TS 1100 °C

 

N

0 TS 1300 °C

 

W

0 TS 2000 °C

 

R.T.D.

Pt100

-200 to 850 °C (FS = 1050 °C)

 

JPt100

-200 to 500 °C (FS = 700 °C)

 

Pt1000

-200 to 500 °C (FS = 700 °C)