HORIBA 8007-10C カドミウムイオン電極

  • 生産者:HORIBA
    Model: 8007-10C
    起源: Japan
    品物を保証します: 3 Month
    • 要件クオート Request a Quotation
  • 連絡先
    • カートへ追加
    • 今すぐ注文
  • 校正検証修理サービスを参照してください

      いい価格のため、連絡しなさい

    ホットライン: (+84) 968060220

    電話番号: +84 (24) 62923267

Measurement Range: 

0.1 to 11,240 mg/L Cd2+

(10-6 to 10-1 mol/L) Cd2+

Temperature Range: 0 - 50 ºC

pH Range: 11 mg/L (10-4 mol/L) Cd2+pH 3 to 8

Applicable temperature range: 0 to 50℃

Response time (90%): Within 10 seconds

Applicable reference electrode: 2060A, 2565A

Selection coefficient:

Cu2+、Hg2+Ag+ = Not acceptable

Pb2+ = 0.1 Fe3+ = 1

Cr3+、Fe2+ = 100

Ni2+ = 1,000


Details

• The response time is the time which is required to reach 90% response when the ion concentration is gradually changed from 10-4 mol/L to 10-2 mol/L with the solution stirred.
Exception:


*1: 90% response when ion concentration is changed to 10-6 mol/L ∼ 10-2 mol/L
*2: 95% response when ion concentration is changed to 10-3 mol/L ∼ 10-1 mol/L
*3: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-2 mol/L
*4: 95% response when ion concentration is changed to 10-4 mol/L ∼ 10-1 mol/L

• The selection coefficient is a ratio of the limit concentration of coexisting ions (mol/L) to the ion concentration to be measured (mol/L); The value of 1000 means that the coexisting ions can be permitted up to 1000 times the ion measured and ''not acceptable'' means that chemical change occurs in the solid response membrane.