For full functionality of this site it is necessary to enable JavaScript.

JEIOTECH JMV-012 加熱及び冷却チャンバー (125L, -35~180℃)

Chamber Volume (L / cu ft)    125 / 4.4
Controller    Programmable, Touch screen type controller (3.5" Color monitor)
Temperature Range (℃ / ℉) :-35~180℃
Temperature Fluctuation  at 25℃ (℃ / ℉)    0.2 / 0.36
Heating Time 2) (min.) from 25 to 180℃    60    
Cooling Time 2) (min.) from 25 to -70℃    100
Heating Rate (℃ / ℉ per min.) from -50℃ to 120℃    2.0
Cooling Rate (℃ / ℉ per min.) from 120℃ to -50℃    2.1
Interior (mm / inch)    500×500×500/ 19.7×19.7×19.7
Exterior (mm / inch)890×1000×1200 mm
Net Weight (kg / lbs)    170 / 374.8
Interface Port    Basically RS-232C [RS-485(Maximum 32 Units Control)-optional]
Electrical Requirements (230V, 50Hz)    1ph / 17.8A , Cat. No. AAHKB012K
 

Details


Constructional Features

  • Heat-resistant silicone double packing completely blocks high heat leakage from the inner chamber. 
  • Two-point door latch also enhances tightness to door closure.
  • Energy efficiency is achieved thanks to optimization of the insulation system.
    - Even if the unit is operating at max. temperature, the surface of the unit doesn’t allow high temperatures of over 51℃. (DIN EN 563) 
  • Reliable design and production of door is guaranteed.
    - Evaluates the results after opening the door by 90 degrees and applying 50kg load to the point of 30cm from hinges.
  • Corrosion resistance stainless steel inner chamber.
  • Space efficient and easy loading chamber design.
    - Included two shelves for loading of large amounts of samples.
    - Heavy load shelving, rated for 40kg are also available. (optional)

Safety Features

  • Over-current & Current Leakage Protection. (ELCB, Earth Leakage Circuit Breaker)
  • Independent Safety Fuse.
    - Important parts such as controller and heaters are highly secured from short circuit. Especially each heater has an individual fuse to completely break a short circuit.
  • Reversed-phase Protection.
    - Electrical connection error indicator helps user to rapidly fix the connection
  • Over-current Protection of Motor.
    - Automatic Shut-off in situations where there is over-current to the Motor.
  • Door Open Warning.
    - Heater & Fan stop when the door is the open state.
  • Electric wiring panel access warning.
    - Automatic stop if the cover of electric wiring panel is opened.

 

Operating Features

  • Suitable for specific use such as heat treatment and drying process.
  • Colorful interactive touch type LCD controller.
  • Set point temperature and present temperature areclearly shown together on display.
  • Sufficient programmable control support.
  • Operating progress can be shown in graph mode for better data understanding.
  • Automatic start/stop can be programmed for unattended operations.
  • User selectable operation mode after power interruption -ether Automatic run or Manual run.
  • Display indicates abnormal temperature changes or power failures.
  • Three different language support. (English / Chinese / Korean)
  • Temperature and other operating parameters can be recorded using a six channel digital recorder.(optional)
  • Programming operation.

    Description No.
    Programmable Pattern Capacity 100
    Pattern Repeat Time 999
    Part repeat time 255
    Max. Segments per a Pattern 100
    Available Max. Segments* 6000
    Programmable Process Time per a Segment 0 to 99 hr 59 min
    * Even though the max. segments per pattern are 100, the total available segments are up to 6000 not 10000, therefore arrangementof the number of segment and patterns will need to be done properly.

 

Software

  • RS-485 communication port(optional) enables connection up to maximum 32 units. Simultaneous monitoring and operating of multiple units is an available feature.
  • Program patterns are simple to set with the use of our software.
  • Easy to check status of real-time operations.
    Graph data supports easy understanding.
  • Storage and documentation of programs.
  • Data with saving capabilities in excel and text format.
  • Storage of program information, backup of value settings, recovery, and temperature values.
  • RS-232 communication port as standard offer.
Model JMV-012 JMV-025 JMV-040 JMV-070 JMV-100
Chamber Volume (L / cu ft) 125 / 4.4 253 / 8.9 420 / 14.8 720 / 25.4 1000 / 35.3
Controller Programmable, Touch screen type controller (3.5" Color monitor)
Temperature Range (℃ / ℉) -35 to 180 / -31 to 356
Temperature Fluctuation 1) 2)
at 25℃ (℃ / ℉)
0.5 / 0.9 0.5 / 0.9 0.3 / 0.54 0.4 / 0.72 0.3 / 0.54
Temperature
Variation 2)
(±℃ / ℉)
at -35℃ 0.7 / 1.26 0.7 / 1.26 0.5 / 0.9 1.0 / 1.8 0.7 / 1.26
at -25℃ 0.7 / 1.26 0.7 / 1.26 0.6 / 1.08 1.1 / 1.98 0.8 / 1.44
at -10℃ 0.8 / 1.44 0.8 / 1.44 0.6 / 1.08 1.2 / 2.16 0.9 / 1.62
at 25℃ 0.8 / 1.44 0.8 / 1.44 0.5 / 0.9 0.8 / 1.44 0.9 / 1.62
at 70℃ 0.9 / 1.62 0.9 / 1.62 0.7 / 1.26 1.2 / 2.16 0.8 / 1.44
at 100℃ 1.0 / 1.8 1.0 / 1.8 0.8 / 1.44 1.2 / 2.16 0.9 / 1.62
at 150℃ 1.2 / 2.16 1.2 / 2.16 1.3 / 2.34 1.1 / 1.98 1.1 / 1.98
at 180℃ 1.4 / 2.52 1.4 / 2.52 1.5 / 2.7 1.3 / 2.34 1.4 / 2.52
Heating Time 2) (min.) from 25 to 180℃ 67 67 46 44 60
Cooling Time 2) (min.) from 25 to -35℃ 48 48 35 35 58
Recovery time 2)
after door was
opened (min.)
at -35℃ 6.0 6.0 6.0 6.5 9.0
at -25℃ 5.0 5.0 6.5 7.0 6.5
at -10℃ 6.5 6.5 6.5 8.5 6.5
at 25℃ 5.5 5.5 4.5 2.0 7.0
at 70℃ 8.5 8.5 8.0 9.5 7.5
at 100℃ 5.5 5.5 4.5 6.0 5.5
at 150℃ 6.5 6.5 7.0 9.0 3.5
at 180℃ 6.5 6.5 6.5 8.0 4.0
  • 良質な取り決め
  • オリジナル保証
  • 宅配便
  • 買い取り簡単化

おまけチャンス‐ニュースを受ける登録