For full functionality of this site it is necessary to enable JavaScript.

Labstac WR321 ダブルステージRO超純水システム

  • 生産者:Labstac
    Model: WR321
    起源: China
    品物を保証します: 12 Month
    • 要件クオート Request a Quotation
  • 連絡先
    • カートへ追加
    • 今すぐ注文
  • 校正検証修理サービスを参照してください

      いい価格のため、連絡しなさい

    ホットライン: (+84) 968060220

    電話番号: +84 (24) 62923267

Water Inlet: Tap water: TDS<200 ppm (Extra pretreatment filter is recommended, if TDS>200 ppm)

Temperature: 5-45°C

Pressure: 1.0-4.0 Kgf/cm²

Flow Procedure**: PF+AC+RO+RO+(UV)+AC+DI+(UF)+TF

Bacteria: <0.1 cfu/ml

Output(25°C)****: 1st stage RO water: 125 L/hr, 2nd stage RO water: 60 L/hr

Pure water outlet: 1st, 2nd stage RO and Ultrapure water

DimensionLxWxH: 760×630×1190 mm

Weight: 80 kg

Standard configuration: Main body (Including 1 set of cartridges) + built-in 2 tank (40L PE tank+2 gallon pressure tank)+ accessory bag

Power Consumption (W): 300 W

Power Supply: AC110-220 V, 50/60 Hz

Note: *The feed water quality will influence the pure water´s quality and cartridges´ life-span. 

**PF:polypropylene spun fiber, AC:active carbon, RO:reverse osmosis, DI:ion exchange, UV:ultraviolet, UF:ultrafiltration, TF:terminal microfiltration. 

***Value of number will be influenced by temperature and feed water quality. 

****All the specifications are tested under the situation:feed water´s TDS=200ppm, 25°C, 50psi and 15% recovery rate.

Resistivity(25°C): 18.2 MΩ.cm

Heavy Metal Ion: <0.1 ppb

TOC***: <3 ppb

Particle (>0.2µm): <1/ml

1st stage RO water's TDS: TDS (ppm, mg/l) < TDS of tap water x 5%

2nd stage RO water's conductivity: 1-5 µs/cm, Organic rejection rate>99% (when MW>200 Dalton), Particles and bacteria rejection rate>99%