For full functionality of this site it is necessary to enable JavaScript.

Memmert ICH750C 湿度チャンバー

  • 生産者:MEMMERT
    Model: ICH750C
    起源: Germany
    品物を保証します: 12 Month
    • 要件クオート Request a Quotation
  • 連絡先
    • カートへ追加
    • 今すぐ注文
  • 校正検証修理サービスを参照してください

      いい価格のため、連絡しなさい

    ホットライン: (+84) 968060220

    電話番号: +84 (24) 62923267

Temperature
Working temperature range     from +10°C up to +50°C
Setting accuracy temperature     0.1°C
Temperature     2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
Humidity
Humidity     humidity supply with distilled water from external tank by self-priming pump
Humidification     humidification by hot steam generator
Humidity     active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
CO2
CO2     CO2-supply via sterile filter
CO2     intrinsic adjustment of altitude above sea level
CO2     CO2-supply interruption upon door opening
CO2     digital electronic CO2-control with automatic zeroing (infrared measuring system) - adjustment range 0 - 10 % - resolution 0.1 %
Control technology
ControlCOCKPIT     TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting     German, English, Spanish, French, Polish, Czech, Hungarian
Timer     Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT     the process time does not start until the set temperature is reached
Calibration     three freely selectable values each, temperature/humidity, for two freely selectable values CO2
adjustable parameters     temperature (Celsius or Fahrenheit), fan speed, CO2, programme time, time zones, summertime/wintertime
Ventilation
Fan     forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps, fixed value 50 % when working with CO2
Communication
Documentation     programme stored in case of power failure
Programming     AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety
Temperature control     mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control     over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY     additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system     integral fault diagnostics for temperature, CO2 and humidity control
Alarm     visual and acoustic
Heating concept
Air jacket     air jacket heating system for gentle all-around heating
Heating concept     no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Cooling     energy-saving use of CFC-free cooling/heating system (refrigerant R134a)
Defrosting     highly efficient automatic defrosting system
Standard equipment
Internals     2 stainless steel grid(s), electropolished
Works calibration certificate     standard values +10 °C and +37 °C, 60 % rh at +30 °C
Door     inner glass doors
Door     fully insulated stainless steel doors with 2-point locking (compression door lock)
Stainless steel interior
Volume     749 l
Dimensions W x H x D in mm     w(A) x h(B) x d(C): 1040 x 1200 x 600 mm
Max. number of internals     14
Max. loading of chamber     200 kg
Max. loading per internal     30 kg
Textured stainless steel casing
Installation     on lockable castors
Dimensions     w(D) x h(E) x d(F): 1224 x 1950 x 834 mm
Housing     rear zinc-plated steel
Electrical data
Voltage Electrical load     230 V, 50/60 Hz
approx. 1350 W
Voltage Electrical load     115 V, 50/60 Hz
approx. 1350 W
Packing/shipping data/Setting Up
Set Up     The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Transport information     The appliances must be transported upright
Customs tariff number     8419 8998
Country of origin     Federal Republic of Germany
WEEE-Reg.-No.     DE 66812464
Dimensions approx incl. carton     B x H x T: 1330 x 2150 x 1050 mm
Net weight     approx. 249 kg
Gross weight carton     approx. 319 kg

Details
  • 良質な取り決め
  • オリジナル保証
  • 宅配便
  • 買い取り簡単化

おまけチャンス‐ニュースを受ける登録