For full functionality of this site it is necessary to enable JavaScript.

Tektronix SA2500 携帯分光アナライザー

Manufacturer: Tektronix
Model: SA2500

Details

Datasheet

Specifications

All specifications apply to all models unless noted otherwise.

Model overview
Model Real-time bandwidth Frequency range Minimum event duration for 100% POI SFDR (typical)
SA2500 20 MHz 10 kHz - 6.2 GHz 125 μs 1 - 500 μs <70 db="" td="">
H500 20 MHz 10 kHz - 6.2 GHz 125 μs <70 db="" td="">

1 With option EP1 installed

General performance characteristics
RF input
Operating frequency range
10 kHz - 6.2 GHz
Maximum operating input level
+20 dBm peak envelope power

This is the maximum input level at which the instrument will meet its performance specifications. For a signal without any amplitude variation, peak envelope power = rms.

Maximum Input Power without Damage
50 Wrms below 3.2 GHz

15 Wrms between 3.2 GHz and 6.2 GHz

IF output
Output impedence
50 Ω
IF center frequency
140 MHz
IF 3 dB bandwidth
24 MHz
IF Output Level (nominal performance at 0 dBm input)
Input frequency IF output level
1 GHz -12 dBm
1.6 GHz -12 dBm
3.6 GHz -10 dBm
4.35 GHz -11 dBm
5 GHz -16 dBm
5.75 GHz -22 dBm
Internal timebase
Frequency error (factory calibration corrected)
±0.5 PPM from 0 °C to 50 °C

±1.0 PPM aging/year

Twenty-minute warm-up period required to meet accuracy specification

Frequency error (GPS corrected)
±0.01 ppm (typical)
Frequency error (after GPS lock loss)
±0.03 ppm, 10 minute interval after lock loss (unit operated for >20 minutes before lock loss and < 5 °C temperature change over interval) (typical)
External reference input
impedance
1500 Ω
Frequencey range
1 MHz up to 20 MHz ± 1 PPM, in 1 MHz steps
Input level range
-15 dBm to +15 dBm, 1 MHz to 15 MHz

-10 dBm to +15 dBm, 16 MHz to 20 MHz

dBm levels assume 50 ohm source

Integrated GPS receiver
Position Accuracy (typical)
Horizontal: R < 9 meters (P = 90%)

Altitude: H < 18 meters (P = 90%)

Position Update Rate (nominal)
1 update/sec (Latitude/Longitude/Altitude)
Spectrum analyzer characteristics
Frequency
Span
10 kHz to 6.2 GHz (preamp off)

10 MHz to 6.2 GHz (preamp on)

Center frequency setting resolution
1 Hz
Swept spans (wide scanning)
Span Range
20 MHz to 6.2 GHz
Spectrums per second
61 (typical)
Scan Speed
1240 MHz per second (typical)
Resolution Bandwidth (RBW)
RBW Range
10 Hz to 3 MHz (Manual)

10 Hz to 1 MHz (Auto)

RBW setting resolution
1 Hz
Spectral purity
Displayed average noise level, preamp On
-153 dBm, 10 MHz to 2 GHz, 10 Hz RBW

-152 dBm, 2 GHz to 4 GHz, 10 Hz RBW

-151 dBm, 4 to 5 GHz, 10 Hz RBW

-145 dBm, 5 to 6.2 GHz, 10 Hz RBW

Phase noise (entire operating frequency range)
≤ -95 dBc/Hz at 10 kHz offset

≤ -95 dBc/Hz at 20 kHz offset

≤ -95 dBc/Hz at 30 kHz offset

≤ -97 dBc/Hz at 100 kHz offset

≤ -110 dBc/Hz at 1 MHz offset

Residual spurious, preamp Off
≤ -90 dBm, 0 dBm attenuator setting

Exception Frequencies:

9 MHz to 19 MHz center frequency

3464 MHz center frequency

4592 MHz center frequency

5374 MHz to 5378 MHz center frequency

6160 MHz center frequency

Residual spurious, preamp On
≤ -105 dBm, 0 dBm attenuator setting

Exception Frequencies:

9 MHz to 19 MHz center frequency

5374 MHz to 5378 MHz center frequency

Third order IMD
≤ -70 dBc for two tones at or below the reference level, preamp Off, all gain settings auto-coupled
Second harmonic
≤ -60 dBc for a single tone at or below the reference level, preamp Off, all gain settings auto-coupled
Input-related spurious
≤ -70 dBc except for Fin= 2.282 GHz ± 20 MHz

The dBc reference for this specification is the total power of all signals at the input of the instrument regardless of the current span

Input-related spurious, exception frequencies, typical
≤ -55 dBc at Fin= 2.282 GHz ± 20 MHz

The dBc reference for this specification is the total power of all signals at the input of the instrument regardless of the current span

Third order intercept
≥ +7 dBm, 0 dB input attenuation, preamp Off
Spectral display amplitude
Reference level range
+20 dBm to -160 dBm
Marker power accuracy
±1.75 dB, -50 dBm ≤ input ≤ +20 dBm, preamp off

±3.0 dB, -80 dBm ≤ input < -50 dBm, preamp on, above 10 MHz

±3.75 dB, -120 dBm ≤ input < -80 dBm, preamp on, above 10 MHz

Use peak detector for CW-like signals; use average detector for wideband (signal >> RBW)

Accuracy guaranteed for CW signals and span set to 20 MHz or less

Display modes
Display modes
Normal – Updates display with each new result

Max Hold – Updates displayed point only if new point > old

Min Hold – Updates displayed point only if new point < old

Max/Min Hold – Displays a vertical bar between Max Hold and Min Hold

Average – Displays average of N (specified by user) acquisitions

Average is calculated as follows:

Last N values are saved in memory; when a new result is available, the earliest result of the N stored values is discarded, the new result is added to the stored values, and a new average is calculated from the stored values. If the number of results is less than N, then all of the results are averaged together

Number of averages
1 ≤ N ≤ 200 
General purpose RF measurement characteristics
RF channel power measurement
Bandwidth range
1 kHz - 20 MHz
Accuracy
≤1.2 dB; +20 dBm to -60 dBm; 1 MHz to 3.2 GHz, preamp off, Ref Level > -35 dBm

≤2.4 dB; -40 dBmto -75 dBm; 10MHz to 3.2 GHz, preamp on, Ref Level ≤ -35 dBm

≤1.8 dB; +20 dBm to -50 dBm; 3.2 GHz to 6.2 GHz, preamp off, Ref Level > -35 dBm

≤3 dB; -40 dBm to -75 dBm; Resolution BW < 100 kHz; -40 dBm to -55 dBm; Resolution BW ≥ 100 kHz 3.2GHz to 6.2 GHz, preamp on, Ref Level ≤ -35 dBm

Specifications apply for default control settings (Auto RBW, Auto Level)

Occupied bandwidth measurement
Percent Power Inclusion Range
50-100%
RF field strength
Channel bandwidth range
Same as RF channel power
Accuracy
Same as RF channel power
Scan, classify, and locate characteristics
DPX™ Live RF spectrum display
Spectrum processing rate, nominal
10,000 spectrums per second, span independent (H500 standard and SA2500 with Option EP1)

2,500 spectrums per second (SA2500 standard)

Minimum signal duration for 100% probability of intercept (POI), typical
125 μs (H500 standard and SA2500 with Option EP1)

500 μs (SA2500 standard)

Span Range
5 kHz to 20 MHz
Spectrogram (rising raster)
Spectrum processing rate, nominal
20 spectrums per second
Minimum signal duration for time measurement
20 milliseconds (typical)
Span range
5 kHz to 20 MHz
Trigger
Modes
Single or continuous, free run or triggered
Event source
IF level, external input, or internal timebase
Types
Rising edge, falling edge, level above threshold, level below threshold
Delay
0 to 60 s with 1 μs resolution
Position
Settable from 0-100%
IF level trigger
Threshold range
-160 dBm to +20 dBm
Bandwidth range
5 kHz to 20 MHz
External trigger
Maximum input level without damage
±5 Vpeak continuous
Minimum high threshold
2.0 V
Maximum low threshold
0.8 V
Minimum High/Low time
10 ns
Impedance
10 kΩ
Coupling
DC
Internal timebase trigger
Mode
Single trigger on time, repeat trigger at interval, or both
Resolution
1 μs
Measurement result time stamps
Resolution (nominal)
1 ms before GPS lock obtained; 1 ns after GPS lock obtained.
Accuracy - relative (typical)
±500 ns time-stamp error between multiple measurement results. Internal GPS reference lock required.
Accuracy - GPS reference (typical)
±1 μs, all measurements except DPX spectrum; ±1 ms DPX spectrum.

Time-stamp error relative to GPS system absolute time reference. Internal GPS reference lock required, identical acquisition bandwidth setting required.

IQ acquisition time

(available in amplitude vs. time measurement)

Span Sample rate Maximum acquisition length
20 MHz 28 msps 36 ms
10 MHz 14 msps 73 ms
5 MHz 7 msps 146 ms
2 MHz 2.8 msps 365 ms
1 MHz 1.4 msps 731 ms
500 kHz 700 ksps 1.4 sec
200 kHz 280 ksps 3.6 sec
100 kHz 140 ksps 7.3 sec
50 kHz 72.9 ksps 14 sec
20 kHz 27.3 ksps 37 sec
10 kHz 13.7 ksps 74 sec
5 kHz 6.8 ksps 149 sec
AM demodulation
Measurement frequency
As selected
Minimum input signal level, typical
-100 dBm
Audio measurement bandwidth
8 kHz
FM demodulation
Measurement frequency
As selected
Minimum input signal level, typical
-100 dBm
Maximum signal deviation
Up to 100 kHz
Audio measurement bandwidth
8 kHz, 15 kHz, 75 kHz, or 200 kHz
Maximum audio output bandwidth
15 kHz
Signal strength indicator
Input signal level
-120 dBm, minimum
Measurement frequency
As selected
Measurement bandwidth
Up to 20 MHz, dependant upon span and RBW setting
Tone type
Variable beep rate or variable frequency
Update rate, typical
10 per second
Mapping
Native map type
Graticule (.gsf)
Map types directly supported
Pitney Bowes MapInfo (*.mif), Bitmap (*.bmp), MPS (collection of GSF maps derived from Open Street Map mapping tiles) 1
Other map types accepted using PC application iMap converter
Google™ earth

Microsoft® MapPoint®

USGS DLG (*.opt)

ESRI ArcInfo Shape (*.shp)

Other raster formats (*.gif, *.jpg, *.png, *.tif)

1 Maps derived from Open Street Map tiles are © OpenStreetMap contributors. For more information see http://www.openstreetmap.org/copyright.

Power
Battery life
5 hours of continuous Spectrum mode (with optional second battery). Actual life can be higher depending on usage.
Physical characteristics
Temperature
Operating
0 °C to +50 °C specified performance, -10 °C to +50 °C, typical
Nonoperating
-40 °C to +60 °C
With batteries installed
While charging: 0 °C to +45 °C

Storage (nonoperating): -20 °C to +60 °C

Humidity
Operating and nonoperating
5% to 95% relative humidity (RH) at up to +30 °C, noncondensing

5% to 45% RH above +30 °C up to +50 °C, noncondensing

Altitude
Operating
Up to 4,600 meters (15,092 feet)
Nonoperating
Up to 12,192 meters (40,000 feet)
Dimensions
Height
25.5 cm. (10.0 in.)
Width
33 cm. (13 in.)
Depth
12.5 cm.
Weight
5.56 kg (12.27 lb.)
Color display
Size
10.4 in. (diagonal), transflective LCD, color
Resolution
640×480 (VGA)
EMC environment and safety
Electromagnetic compatibility (EMC) compliance
EN61326-1:2006 and EN61326-2:2006 Product Family Standard for Electrical Equipment for Measurement, Control, and Laboratory Use – EMC Requirements.
European Union
Emissions
CISPR11, group 1, class a

EN 61000-3-2 

EN 61000-3-3 

Immunity
IEC 61000-4-2 

IEC 61000-4-3 

IEC 61000-4-4 

IEC 61000-4-5 

IEC 61000-4-6 

IEC 61000-4-11 

Australia/New Zealand
EMC compliance in accordance with the ACMA
USA
FCC, CFR Title 47, Part 15, Subpart B, Class A.
Safety compliance
ANSI/UL610101:2004 

Electrical Equipment for Measurement, Control, and Laboratory Use

CSA C22.2 No. 61010.1:2004 

Electrical Equipment for Measurement, Control, and Laboratory Use

EN 610101:2001 

Safety Compliance Electrical Equipment for Measurement, Control, and Laboratory Use

IEC610101:2001 

Electrical Equipment for Measurement, Control, and Laboratory Use

ISA 82.02.01 

Electrical Equipment for Measurement, Control, and Laboratory Use

 

  • 良質な取り決め