YuYang YY503 직업적인 단 하나 불타는 품목 시험 기구/SBI 불 시험 장비

Operating Conditions:

1. Power supply: AC 220V 50HZ;

2. Power: 8.5KW;

3. Ambient temperature: 5~35 ℃;

4. Relative humidity:20%~80%;

5. Atmospheric pressure:0~250KPa.


Specifications:

1. Test room: L3000*W3000*H2400(MM),at the top of room connecting hood with sampling pipe and exhaust ducting,there is space for natural gas in-out blew the trolley.During testing,the combustion heat release of sample and combustion production can be discharged from exhaust duct.With brick building.

2. Ignition source,placed on trolley vertical rectangular box in the corner of the 31 kw propane sandbox burners(two equal sides 0f 250mm,height 80mm);

3. Exhaust pipe is provided with general measurement section for placing sensors and sampling tubes;

4. Exhaust speed: 0.50 m3 /S~0.65 m 3/ S;

5. Test room ambient temperature measurement:K-type sheathed thermocouples with diameter 2mm,ambient pressure test accuracy: ± 200Pa;

6. Flue temperature measurement:K-type sheathed thermocouples with diameter 0.5mm;

7. Temperature measurement accuracy: 0.5 ℃;

8. Flue high-precision pressure sensor accuracy: ± 2Pa;

9. Time recording system accuracy: 0.1S;

10. Test time: 1 to 30 minutes can be set;

11. Smoke density measurement system to measure the density of the smoke during testing;

12. Gas analyzer, and the key components are imported.

13. Embedded computers and 8.4-inch LCD display with TCP / IP and RS-232 communication interfaces can be choice;

14. Oxygen measurement:

1 ) Measuring range: 0-25%

2) Signal output: 4-20mA;

3 ) Response time T90: ≤2S;

4 ) Ambient temperature: 0-45 ℃;

5 ) Relative humidity: <90% (non-condensing);

6 ) Linearity: <± 0.1% O2;

7 ) Zero drift: 0.05% O2 per week;

8 ) Repeatability: <± 0.02% O2;

15. Carbon dioxide(CO2) measurement:

1) Measuring range:0-10%;

2) Repeatability: <± 1% internal structure of the control cabinet;

3) Zero drift: ≤2% / week;

4) Span drift: ≤2% / week;

5) Linearity deviation: <± 1%;

6) Response time: T90≤2 seconds;

16. Record data acquisition system can collect oxygen concentration, carbon monoxide concentration, temperature, smoke density, heat release rate, mass loss rate, etc., can be stored.


Details

Datasheet