Digital MultiMeter Calibration Procedure

06/04/y 10:26:04

**1.Range of application** \- DC Voltage 1000V \- AC Voltage 1000V, Frequency 100kHz \- DC Current 30A \- AC Current 30A, Frequency 10kHz \- Resistor 100

1.Range of application

- DC Voltage 1000V

- AC Voltage 1000V, Frequency 100kHz

- DC Current 30A

- AC Current 30A, Frequency 10kHz

- Resistor 1000MΩ

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s): Multifunction Calibrator (MFC)

Meatest M142 Multifunction Calibrator 15 ppm

- DC Voltage: 0µV - 1000V.

- AC Voltage: 0,1mV - 1000V, Frequency: 20Hz - 100kHz.

- DC Current: 0µA - 30A.

- AC Current: 1µA - 30A, Frequency: 20Hz - 10kHz.

- Resistor: 0 Ω - 1GΩ.

3.2 Auxiliary Instrument(s)

3.2.1 Testo Hygro-Thermometer 608-H1

Testo 608-H1 - Thermal Hygrometer (Part Number 0560 6081) | TEquipment

- Range of Measurement: (0 ÷ 50) °C; Accuracy: ±0,5 °C

- Range of Measurement: (10 ÷ 95) %RH; Accuracy: ±3 %RH

3.2.2 Dedicated Measuring Probe

- Probes with small electromotive force.

4. Conditions for Calibration

- Temperature: (23 ± 5) °C.

- Humidity: no higher than 80% RH.

5. Prepare for Calibration

- Make sure the power supply to the DMM meets the requirements specified in the catalog, the fuses, the protective circuit are still working properly.

- The DMM should be placed in the calibration environment for at least 8 hours.

- MFC and DMM must be turned on and dried for at least 30 minutes prior to calibration.

6. Begin the calibration

a. Outside Check

b. Technical Check

c. Measurement Check

    1. General requirements

    2. Begin the DMM Calibration

        a. DC Voltage

        b. AC Voltage

        c. DC Current

        d. AC Current

        e. Resistor

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated DMM is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of DMM is 12 months.


For calibration of DMM, click here.

For further information about Digital MCalibration service, please contact

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220


ຂ່າວທີ່ກ່ຽວຂ້ອງ

How do HRC, HRB, and HRA scales differ in hardness testing?
15/06/y 14:30:32

HRC, HRB, and HRA all belong to the Rockwell hardness system but are applied to different material groups. Understanding the differences between HRC, HRB, and HRA helps in selecting the correct testing method, accurately reading specifications, and avoiding unnecessary errors in material quality assessment.

Common Errors in Fiber Loss Optic
13/06/y 09:24:06

Optical power meter used together optical light source, can be applied to determine optical fiber link attenuation — a parameter that reflects the overall quality of the fiber optic cabling system

ຕິດຕາມຂ່າວສານ ແລະ ຂໍ້ສະເໜີ

ຮັບສ່ວນຫຼຸດພິເສດຕາມປະລິມານ, ອັບເດດລາຄາຂາຍສົ່ງ ແລະ ການແຈ້ງເຕືອນສິນຄ້າໃໝ່ສົ່ງກົງເຖິງອິນບັອກຂອງທ່ານ.

ໂດຍການສະໝັກສະມາຊິກ, ທ່ານຍອມຮັບ ເງື່ອນໄຂການໃຫ້ບໍລິການ ແລະ ນະໂຍບາຍຄວາມເປັນສ່ວນຕົວ ຂອງພວກເຮົາ.

ການຊ່ວຍເຫຼືໍາດ່ວນ

ເຂົ້າເຖິງຜູ້ຊ່ຽວຊານທີ່ໄດ້ຮັບການຢັ້ງຢືນຂອງພວກເຮົາໂດຍກົງ