| Input A and input B |
| DC voltage (VDC) |
| Ranges | 500 mV, 5 V, 50 V, 500 V, 750 V |
| Accuracy | ±(0.5% +5 counts) |
| Common mode rejection (CMRR) | >100 dB @ dc, >60 dB @ 50, 60, or 400 Hz |
| Full scale reading | 5000 counts |
| True-rms voltages (V ac and V ac+dc) |
| Ranges | 500 mV, 5 V, 50 V, 500 V, 750 V |
| Accuracy for 5% to 100% of range (DC coupled) | DC to 60 Hz (V ac+dc) | ±(1% +10 counts) |
| 1 Hz to 60 Hz (V ac) | ±(1% +10 counts) |
| Accuracy for 5% to 100% of range (AC or dc coupled) | 60 Hz to 20 kHz | ±(2.5% +15 counts) |
| DC rejection (only VAC) | >50 dB |
| Common mode rejection (CMRR) | >100 dB @ dc |
| >60 dB @ 50, 60, or 400 Hz |
| Full scale reading | 5000 counts, reading is independent of any signal crest factor. |
| Peak |
| Modes | Max peak, Min peak, or pk-to-pk |
| Ranges | 500 mV, 5 V, 50 V, 500 V, 2200 V |
| Accuracy | Accuracy Max peak or Min peak | 5% of full scale |
| Accuracy Peak-to-Peak | 10% of full scale |
| Full scale reading | 500 counts |
| Frequency (Hz) |
| Ranges | 123B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz |
| 124B and 125B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz, 10 MHz, and 70 MHz |
| Frequency range | 15 Hz (1 Hz) to 50 MHz in continuous autoset |
| Accuracy @1 Hz to 1 MHz | ±(0.5% +2 counts) |
| Full scale reading | 10,000 counts |
| RPM |
| Max reading | 50.00 kRPM |
| Accuracy | ±(0.5% +2 counts) |
| Duty cycle (PULSE) |
| Range | 2% to 98% |
| Frequency range | 15 Hz (1 Hz) to 30 MHz in continuous autoset |
| Pulse width (PULSE) |
| Frequency range | 15 Hz (1 Hz) to 30 MHz in continuous autoset |
| Full scale reading | 1000 counts |
| Amperes (AMP) |
| With current clamp | Ranges | Same as V dc, V ac, V ac+dc, or PEAK |
| Scale factors | 0.1 mV/A, 1 mV/A, 10 mV/A, 100 mV/A, 400 mV/A, 1 V/A, 10 mV/mA |
| Accuracy | Same as V dc, V ac, V ac+dc, or PEAK (add current clamp uncertainty) |
| Temperature (TEMP) with optional temperature probe |
| Range | 200 °C/div (200 °F/div) |
| Scale factor | 1 mV/°C and 1 mV/°F |
| Accuracy | As V dc (add temp. probe uncertainty) |
| Decibel (dB) |
| 0 dBV | 1 V |
| 0 dBm (600 Ω /50 Ω) | 1 mW referenced to 600 Ω or 50 Ω |
| dB on | V dc, V ac, or Vac+dc |
| Full scale reading | 1000 counts |
| Crest factor (CREST) |
| Range | 1 to 10 |
| Full scale reading | 90 Counts |
| Phase |
| Modes | A to B, B to A |
| Range | 0 to 359 degrees |
| Resolution | 1 degree |
| Power (125B only) |
| Configurations | 1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only) |
| Power factor (PF) | Ratio between watts and VA range - 0.00 to 1.00 |
| Watt | RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes) |
| Full scale reading | 999 counts |
| VA | Vrms x Arms |
| Full scale reading | 999 counts |
| VA reactive (var) | √((VA)2-W2) |
| Full scale reading | 999 counts |
| Vpwm |
| Purpose | To measure on pulse width modulated signals, like motor drive inverter outputs |
| Principle | Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency |
| Accuracy | As Vrms for sinewave signals |