Indicator: 10 ~ 200 mg/L
Measuring Range: 10 ~ 200 mg/L
Sampling Time:
Colour Change:
Reaction Principle:
Nr. of tests: 10
Catalog
Performance
| Measuring Range |
10 to 200 mg/L |
| Sampling Time |
1 minutes |
| Detecting Limit |
3 mg/L |
| Colour Change |
Brown → White |
| Reaction Principle |
2Cl- + Ag2CrO4 → 2AgCl + CrO42- |
| Shelf Life |
3 Years |
| Corrections for temperature & humidity |
Unnecessary |
| Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
| Substance |
Concentration |
Interference |
Changes colour by itself to |
| CN- |
≧ 10 mg/L |
+ |
White ( ≧ 1mg/L) |
| Br- |
≧ 10 mg/L |
+ |
White ( ≧ 10mg/L) |
| SCN- |
≧ 1 mg/L |
Unreadable |
White ( ≧ 100mg/L) |
| Fe2+ |
≧ 20 mg/L |
 |
No ( ≦ 1000mg/L) |
| Fe3+ |
≧ 20 mg/L |
+ |
White ( ≧ 20mg/L) |
| F- |
≦ 1000 mg/L |
No |
No( ≦ 1000mg/L) |
| I- |
≧ 20 mg/L |
+ |
Pale yellow ( ≧ 20mg/L) |
| S2- |
≧ 3 mg/L |
+ |
Blackish gray ( ≧ 3mg/L) |
| SO42- |
≦ 1000 mg/L |
No |
No ( ≦ 1000mg/L) |
Other substance measurable with this detector tube
| Substance |
Correction Factor |
Measuring Range |
Detector tubes for dissolved substances
(i.e. in solution).
Dissolved sulphide detector tube
211H / 211M / 211 / 211LL

|
 |
Liquid detector tube
| Ozone: |
218 |
| Chloride ion: |
221L / 221LL |
| Mercury: |
271 |
| Hexavalent chromium: |
273 |
| Bivalent iron: |
281 |
| Copper: |
284 |
| Zinc: |
285 |
| Nicke: |
291 |
|
To measure, simply immerse the detector tube in a sample of liquid (placed in a beaker or other container). Anyone can measure concentrations quickly, easily and accurately on site.