Probe: FM (included) magnetic induction probe
Range: 0-1500μm
Resolution: 0.1μm (<100μm)
Accuracy*:
Zero calibration: ±(1μm+2%L)
One-point calibration: ±(1μm+2%L)
Two-point calibration: ±(1μm+(1-2)%L)
Multi-point calibration: ±(1μm+1%L)
Measuring mode: single point measurement, scan mode, differential mode, average mode
Calibration mode: zero calibration, one-point calibration, two-point calibration, multi-point calibration
Test frequency: 3 times per second
Minimum substrate thickness: 0.5mm
Minimum measuring area: Ø7mm
Minimum curvature radius of convex workpiece: 1.5mm
Data storage: 500 groups
Interface: USB
Operation temperature: -10°C~50°C
Power supply: 3x1.5V pin AAA
Dimension: 150×70×30mm
Weight: 160g
Probe: NM (optional) eddy current probe
Range: 0-1500μm (chrome plating on copper: 0-40μm)
Resolution: 1μm (100μm~1500μm)
Accuracy*:
Zero calibration: ±(1μm+2%L)
One-point calibration: ±(1μm+2%L)
Two-point calibration: ±(1μm+(1-2)%L)
Multi-point calibration: ±(1μm+1%L)
Measuring mode: single point measurement, scan mode, differential mode, average mode
Calibration mode: zero calibration, one-point calibration, two-point calibration, multi-point calibration
Test frequency: 3 times per second
Minimum substrate thickness: 0.3mm
Minimum measuring area: Ø5mm
Minimum curvature radius of convex workpiece: 3mm
Data storage: 500 groups
Interface: USB
Operation temperature: -10°C~50°C
Power supply: 3x1.5V pin AAA
Dimension: 150×70×30mm
Weight: 160g
*L is the measured value in µm.
Standard Delivery:
Main unit: 1pc
Magnetic induction preobe (FM): 1pc
Zero calibration block for FM probe: 1pc
Calibration foils (12/50/100/250/500/1000μm): 6pcs
AAA battery: 3pc
Software and USB cable: 1pc
- Magnetic induction probe (FM) measures the thickness of non-magnetic coating and non-metallic coating on magnetic metal substrate.
Substrate: steel, iron, alloy, hard magnetic steel, etc.
Coating: zinc, aluminum, chrome, copper, rubber, paint, etc.
- Eddy current probe (NM) measures the thickness of non-conductive coating on non-magnetic metal substrate.
Substrate: copper, aluminumn, zinc, tin, etc.
Coating: rubber, paint, plastic, anodized film, etc.
- Real-time temperature compensation guarantees high accuracy, thin plating and
oxide layer less than 20μm can be measured accurately
- Reduces the effects of electromagnetic interference and hand-held operation Probe can be re-matched after abrasion
- Tolerance measurement with adjustable alarm threshold
USB interface for data transmission
Coupling status indication
- Support cable printer