Volume: 241 liters
Interior dimension: width 600 x height 810 x depth 500 mm
Exterior dimension: width 759 x height 1180 x depth 691 mm
Stainless steel shelves (standard): 2
Max. number of perforated shelves: 12
Working-temperature range: +5°C above ambient temperature up to 50°C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting accuracy: 0.1°C
2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature fluctuations with time at 37 °C +/- 0.1 K
Temperature variation in chamber at + 37 °C +/- 0.3 K
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93% rh ± 2.5 %
Setting range active humidity control: 40 to 97%rh and rh-off (with option K7)
Setting accuracy humidity: 0.5% rh
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO2: 0 - 20%
Variation in time CO2: ± 0.2%
Setting accuracy CO2: 0.1%
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour display
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Function SetpointWAIT: the process time does not start until the set temperature is reached
Adjustable parameters: temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime
Interface: Ethernet LAN, USB
Programme stored in case of power failure
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Additionally integrated over and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system: integral fault diagnostics for temperature, CO2
Alarm: visual and acoustic
Large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
Fully insulated stainless steel door with 2-point locking
Inner glass door with opening (8 mm Ø) to take gas sample
Interior: material 1.4301 (ASTM 304), corrosion resistant
Voltage: 230V, 50/60Hz
Electrical load: aprrox. 1650 W
Ambient temperature: 10 – 35°C
Humidity: max. 70%, non-condensing