Working-temperature range:from -12°C up to +60°C (Optimum performance of cooling aggregate at an ambient temperature of +16°C up to +34°C. Not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over.)
Setting accuracy temperature0.1 °C
Temperature:2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
ControlCOCKPIT:TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting:German, English, Spanish, French, Polish, Czech, Hungarian
Timer:Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT:the process time does not start until the set temperature is reached
Calibration:three freely selectable temperature values
adjustable parameters:temperature (Celsius or Fahrenheit), fan speed, programme time, time zones, summertime/wintertime
Documentation:programme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Temperature control:mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control:over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY:additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system:for fault analysis
Alarm:visual and acoustic
Cooling:Cooling compressor with climate-neutral refrigerant CO2 (R744), GWP1
Air jacket:air jacket heating system for gentle all-around heating
Heating conceptno drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Defrosting:highly efficient automatic defrosting system
Door:fully insulated stainless steel door with 2-point locking (compression door lock)
Internals:2 stainless steel grid(s), electropolished
Works calibration certificate:for +10°C and +37°C
Door:inner glass door
Stainless steel interior
Dimensions:w(A) x h(B) x d(C): 640 x 800 x 500 mm (d less 33 mm for fan)
Volume:256 l
Max. number of internals:9
Max. loading of chamber:200 kg
Max. loading per internal:20 kg
Dimensions:w(D) x h(E) x d(F): 824 x 1552 x 684 mm (d +56mm door handle)
Installation:on lockable castors
Housing:rear zinc-plated steel
Voltage Electrical load:230 V, 50 Hz approx. 1200 W
Voltage Electrical load:115 V, 50/60 Hz approx. 1200 W
Dimensions approx incl. carton:w x h x d: 930 x 1760 x 930 mm
Net weight:approx. 162 kg
Gross weight carton:approx. 219 kg