直接联系认证专家
Gamma radiation detector: Scintillator NaI(Tl) Ø63x63 mm
Energy range: 50 keV – 3 MeV
Measurement ranges (2π geometry):
- Surface activity of 134Cs and 137Cs: 4 – 3700 kBq/m2 (0.1 – 100 Ci/km2)
- Specific activity of 134Cs and 137Cs by in situ method: 50 – 10^6 Bq/kg
- Specific effective activity of 40K, 226Ra and 232Th: 100 – 10^4 Bq/kg
Measurement ranges (4π geometry):
- Specific activity of 134Cs and 137Cs: 50 – 10^6 Bq/kg
- Specific activity of 131I: 30 – 10^6 Bq/kg
- Specific effective activity of 40K, 226Ra and 232Th: 50 – 10^4 Bq/kg
Limit of intrinsic relative error of activity measurement: ±20%
Typical resolution at 662 keV (137Cs): 8%
Typical sensitivity to gamma radiation:
241Am: 11600cps/(µSv·h^-1 )
137Cs: 2200cps/(µSv·h^-1 )
60Co: 1200cps/(µSv·h ^-1 )
Measurement range of ambient dose equivalent rate: 0.03 – 130 µSv/h
Typical sensitivity to 137Cs gamma radiation: 2200 cps/(µSv·h^-1)
Protection class: IР67
Operating temperature range: -20°C to+50°C
Relative humidity with air temperature ≤35°C without condensation: ≤95%
Overall dimensions, weight: Ø130x500 mm, 4.5 kg