直接联系认证专家
Geometry: 45/0 (45 right-shaped illumination, 0 degree viewing)
Light Source: CLEDs
Sensor: Dual Optical Path Array Sensor
Non-contact Measurement Distance: 7.5mm ± 3mm (repeat-ability dE <0.1 for distance fluctuations)
Calibration Method: Built-in Calibration Tile
Wavelength Range: 400nm-700nm
Half Band Width: 5nm
Wavelength interval: 10nm
Non-contact Measurement Distance: 7.5mm ± 3mm (repetitive dE <0.1 for distance fluctuations)
Measuring Aperture: φ7mm
Measure Time: Minimum 20ms
Measurement Interval: 1s
Reflectance Range: R(0%-200%)
Resolution: 0.0001
Measurement parameter: CIE-L*a*b, L*C*h, L*u*v, XYZ, Yxy, reflectivity WI (ASTM E313-00, ASTM E313-73, CIE/ISO, AATCC, Hunter, Taube Berger Stensby) , YI (ASTM D1925, ASTM E313-00, ASTM E313-73), Tint (ASTM E313, CIE, Ganz), metamerism index Milm, color fastness, color fastness
Observer: 10°/2°
Illuminants: A,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30.DLF,NBF,TL83,TL84.U35
Color Difference Formula: ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*94,ΔE*00
Repeatability (standard white tile): Chromaticity value: ΔE*ab standard deviation ≤ 0.03, maximum value ≤ 0.05.
(After calibration, the white calibration tile is measured 30 times at 1s interval)
Inter-instrument Agreement: Within ΔE*ab 0.2 (average measurement of BCRA Series II 12 tiles)
Light Source Life: Lifetime Warranty
Working Temperature: 0°C-45°C, relative humidity 80% or lower (at 35°C), no condensation
Storage Temperature: -25°C-55°C, relative humidity 80% or lower (at 35°C), no condensation
Interface: Industrial USB (others can be customized)
Power Supply: +12DCV
Weight: 2.3kg
Size: 154mm*113mm*145mm (L x W x H)