直接联系认证专家

Frequency
Range: 1 MHz ~ 500 MHz
Resolutionz: 1 kHz
Oscillator level
Cable length = 1 m
Power range (When 50 Ω LOAD isconnected to test port): –40 dBm to 1 dBm
Current range (When SHORT is connected to test port): 0.0894 mArms to 10 mArms
Voltage range (When OPEN is connected to test port): 4.47 mVrms to 502 mVrms
Uncertainty (When 50 Ω LOAD is connected to test port)
(23 °C ± 5 °C)
± 2 dB (frequency ≤ 1 GHz)
± 3 dB (frequency > 1 GHz)
(5 °C to 40 °C)
± 4 dB (frequency ≤ 1 GHz)
± 5 dB (frequency > 1 GHz)
Resolution 0.1 dB (When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.)
Cable length = 2 m (When option 002 is used)
Power range Subtract the following attenuation from the power
Dimensions, weight:
Weight Main unit: 13 kg, test head: 250 g with plate