直接联系认证专家
Division value (d): 5mg
Maximum range (Max): 51000g
Repeatability (5% loading)
+ Standard deviation: 7.5mg
+ Typical error: ±15mg
Repeatability (Close to 100% load)
+ Standard deviation: 10mg
+ Typical error: ±25mg
Linear error
+ Standard deviation: 7.5mg
+Typical error: ±15mg
Position error (Test location according to OIMLR76 specification)
+ Test Weight: 20000g
+ Standard deviation: 7.5mg
+ Typical error: ±15mg
Sensitivity drift (+10℃~+30℃): 5PPM
Maximum peeling range: Less than 100% of the maximum range
Accuracy class: Class I
Verification index value (e) (According to 2014/31/EU specification): 50mg
Typical stability time: 10s
Typical measurement time: 12s
External calibration weight: 50kg, 20kg
Standard weight grade: E2(accordance with OIML R111-1)
Pan Size: ϕ200mm