Test index: Moisture, ash ,volatile matter and LOI
Continuous test in batches: Yes
Standards: ISO 17246, ISO 18123/18122/18134, ISO 11722, ISO 1171,ASTM D3173/D3174/D3175/D7582, GB/T 30732, GB/T 212
Precision: 0.02% RSD
Furnace temperature: Room temperature~1000℃
Temperature control precision: ±2℃
Temperature resolution: 0.1℃
Reproducibility: Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%)Volatile matter: 0.50% (<20.00%); 1.00% (20.00~40.00%); 1.50% (>40.00%)
Cooling time required for batches of test: No
Furnace Cooling (forced air/N2): 950 to 500°C<30 min
Balance Resolution: 0.0001g
Sample introduction and taking: Automatic
Max. Samples: 19 samples + 1 blank sample position
Quantity of sample / test time:: Full indexes for 19 samples< 120 min (Quick method)
Sample weight: 0.5 up to 5g
Max. power: 4kW
Interface:: RS232
Furnace Heating: Resistance Wound
Temperature Calibration: Standard thermocouple
Temp. control precision: ±2℃ (moisture) ±5℃ (ash, volatile matter)
Gas Flow Rate: GB:Moisture:Nitrogen,30~60 times/h Ash:Air 60~120 times/h ASTM:Moisture:Nitrogen or Air 2~4(Furnace Volume /min)volatile matter:Nitrogen 2~4(Furnace Volume /min)
Power requirement: AC 220 V(±10%),50 / 60Hz
Gas requirement: Nitrogen/Oxygen/Compressed Air
Gas Pressure: Air 5 to 6 bar (75 to 90 psi)
Minimum Gas Purity: Air/Nitrogen/Oxygen 99.5%; dry, oil-free
Sample weighing: Under normal temp.
Test Temperature: Moisture 105℃,Ash 815℃,Volatile matter 920℃~950℃ Can be customized
Humidity: <80%
Dimensions: 550x590x530mm
Weight: 70kg
Operational Control: Windows®-based software on external PC/Windows7/ WindowsXP
Advance Calculation: Yes
Analysis Report: Yes
Statistics Report: Yes
System Monitor: Yes