IETLAB PRS系列可编程十进制电阻箱(1mΩ 至100MΩ)

(Full specifications can be found on datasheet)

IEEE, RS232 or Ethernet Programmable Decade Resistor
Wide range and resolution: 1 mΩ to 100 MΩ
Available at accuracies of : 0.01%, 0.02%, 0.05%, 0.1%, 0.5%, and 1%
Special RTD and custom configurations
Dual / Combo rackmount models ( examples: resistance & capacitance )
Consider NEW PRS-300 Decade Resistor
IET PRS20x_LabView Driver for IEEE-488.2 and RS-232 Versions
IET PRS Ethernet_LabView Driver for Ethernet Version
effective programmable impedance unit controlled manually and by a computer
High power versions available
Programmable "open circuit" and "short circuit" states optional
Desktop and 19" Rackmount enclosures
Contact us for Custom or Built to Order Models

Details

datasheet

manual

BASIC SPECIFICATIONS

(See datasheet for full specifications)

Standard Models

Model
PRS-200
PRS-201
PRS-200W
PRS-201W
PRS-202
PRS-202W
PRTD-X-6-0.001
Type
Resistance
Precision Resistance
Wide Range Resistance
Wide Range Precision Resistance
High Precision Resistance
Wide Range High Precision Resistance
Programmable RTD Simulator
Accuracy
1% +70 mΩ
0.1%+ 30 mΩ
1%+ 70 mΩ
0.1%+ 30 mΩ
0.05% + 15 mΩ
0.05% + 15 mΩ
0.01% + 15 mΩ
Decades
7
7
9
9
7
9
6
Range
0 - 10 MΩ
 0 - 10 MΩ
0 - 100 MΩ
 0 - 100 MΩ
0 - 10 MΩ
 0 - 100 MΩ
 4-1,003.99 Ω
Resolution
 1 Ω
 1 Ω
 0.1 Ω
 0.1 Ω
 1 Ω
 0.1 Ω
0.001 Ω
Type of Components
Resistance wire for 0.1 Ω steps and under; metal film for 1 Ω steps and over.
Resistance wire for 0.1 Ω steps and under; wirewound, sealed non-inductive resistors for 1 Ω to 100 kΩ steps; Precision metal-oxide film resistors for 1 MΩ steps and over
Max. Load*
 0.5 A, 200 V (dc + ac peak), 0.2 W/step, 2 W unit, whichever applies first.*
3 A, 200 V (dc + ac peak), 0.5 W/step, 4.5W/unit, whichever applies first.*
Residual Impedance
<450 mΩ
 <600 mΩ
 <100 mΩ
<140 mΩ (typically <100 mΩ)
Absolute Value
* These specifications are dynamic switching limits. The maximum voltage, power, or current which may be applied at any particular resistance setting may be higher as long as the setting is unchanged, or the unit is switched dry.

Switching time: < 4 ms per change; < 7 ms for ≤0.05% units
IEEE: GPIB standard 24 pin connector, conforms to IEEE-488.2; SCPI 1994.0 command set; Hardware or software configurable addressing range of 0 to 30. .
RS232: 25 pin male DTE interface conforms to EIA-STD-RS-530; SCPI 1994.0 command set; data rates from 300 to 115200 bps.
Ethernet: Interface conforms to IEEE 802.3 Speeds 10BaseT (10 Mb/s) and 100BaseT (100 Mb/s), IP Address Static or DHCP, Factory setting 192.168.1.254 static.
A broad range of programmable resistance decade boxes for applications requiring a programmable impedance unit, controlled manually and by a computer via an IEEE-488, RS-232 or Ethernet interface.


    The PRS Series is a broad line of high programmable decade boxes. They provide direct resistance substitution as well as RTD (Resistance Temperature Detector) simulation, in a wide selection of ranges, tolerances and ratings.

    The PRS programmable resistance decade box is a precision resistance source with excellent characteristics of stability, temperature coefficient, and power coefficient. High dynamic ranges are available, starting as low as 1 mΩ, and extending to as many as 10 decades. These features combined with a low virtually constant zero resistance make for very versatile instruments.

    The PRS Series programmable resistance decade box features two optional special settings. An "open circuit" and a "short circuit". These modes are useful for obtaining reproducible transitions between settings, i.e. break-before-make or to short between settings. The "short circuit": setting also provides a reduced zero resistance.

    Operation is both local using convenient direct-reading front panel thumbwheel switches, and remote with optional Ethernet, RS-232, or IEEE-488 interfaces.
    Both can provide an optional extra "10" position for each decade.

    The PRS Series programmable resistance decade box employs very low resistance, low thermal emf relays with gold-clad silver-alloy contacts. A special design keeps contact resistance to a minimum. The gold plating keeps the silver contacts from becoming tarnished when unused, or when only low currents are passed through them. This is most often the case when only minute test currents are drawn by digital multimeters and other test instruments. Contact resistance remains low and repeatable.

    The programmable decade box employees, high-quality gold-plated tellurium-copper five-way binding posts serve to minimize the thermal emf effects, which would produce errors in dc resistance measurements. All other conductors within the instrument, as well as the solder employed contain no metals or junctions that could contribute to thermal emf problems.

    With a resolution as low as 1 mΩ and a maximum available resistance of over 100 MΩ, the PRS Series Series may be used for exacting precision measurement
    applications requiring high accuracy, good stability, and low zero resistance. They are suited for automatic and manual calibration and testing, simulation of RTD'S, programmable loads, and many other laboratory and industrial applications. The PRS Series programmable resistance decade box may be rack mounted to serve as components in measurement and control systems.

    PRTD Version: Low resistance versions with a fixed minimum resistance setting (4 Ω or specified by customer) are suitable for Programmable RTD (Resistance Temperature Detector) simulations. This design virtually eliminates the effect of zero resistance and relay contact resistance, providing the specified absolute accuracy over its entire range.

    DRIVERS - TEST KEYBOARD

    To aid the user in operating the PRS/PCS/PLS, Three “Keyboard” Controller programs - the easiest way to control GPIB, RS232 or Ethernet devices without writing a program - are available from IET. This Keyboard programs automatically finds your device at start-up and it lets you enter just the data that you want to send to the device. This program works with ICS, Measurement Computing and National Instruments GPIB controllers. 

    For GPIB   IEEE-488.1 and IEEE-488.2 interface