Measuring range and Accuracy (at 23°C±5°C)
Range/ Measuring range/ Resolution/ Measuring current/ Accuracy (SLOW)/ Accuracy(FAST)
20mΩ/ 0.00 ~19.99mΩ/ 10μΩ/ 5mA/ ±(0.5%+2digit)/ ±(0.5%+5digit)
200mΩ/ 0.0 ~199.9mΩ/ 100μΩ/ 500μA/ ±(0.3%+2digit)/ ±(0.3%+3digit)
2Ω/ 0.000 ~ 1.999Ω/ 1mΩ/ 50μA/ ±(0.2%+1digit)/ ±(0.2%+2digit)
20Ω/ 0.00 ~ 19.99Ω/ 10mΩ/ 50μA/ ±(0.2%+1digit)/ ±(0.2%+2digit)
200Ω/ 0.0 ~ 199.9Ω/ 100mΩ/ 5μA/ ±(0.2%+1digit)/ ±(0.2%+2digit)
Test signal: 1kHz±0.1% sine wave/20mV(MAX)
Measuring method: 4-terminal measurement [contact check]
Sampling time FAST:40msec. (approx.)/SLOW:140msec. (approx.)
[In case of measurement with contact check function, measuring time is longer by about 5 msec.]
Zero compensation: 50 counts in each range for both
Measuring terminals DC max. voltage: DC 50V
Range changeover: Auto Manual (20mΩ~200Ω)
Indication of comparator's comparison result: LED indication LO/GO/HI and buzzer
Control signal
Remote start signal
Remote hold signal
comparison output [LO/GO/HI]:open collector
contact error output [CE]:open collector
end of comparison output [EOC]:open collector
Operation condition [Temp.]: +5*C~+40*C [Humidity] less than 85%
Power supply: AC100V~240V selectable, 50/60Hz, less than 50VA
Outer dimension: about 250 (W) ×112 (H) ×335 (D) mm (excluding protruding parts such as rubber legs, etc.)
Weight: about 3.0kg