Viscosity range: 36,000 mPa·s at ambient and measuring temperature
Supported syringes:
2 mL Luer plastic for sample viscosity up to 36,000 mPa·s
5 mL Luer plastic for sample viscosity up to 10,000 mPa·s
10 mL Luer plastic for sample viscosity up to 9,000 mPa·s
5 mL glass for sample viscosity up to 10,000 mPa·s
Filling mode: Force applied by syringe
Minimum sample volume per measurement: approx. 3 mL depending on system configuration
Rinsing liquid consumption: approx. 7 mL (if rinse time: 10 s) approx. 11 mL (if rinse time: 15 s)
Compressed air supply:
0.3 bar (4.35 psi) relative (air supply of the master instrument)
max. 1.5 bar (21.8 psi) relative (air supply of Xsample 340)
optional: 2.0 bar (29 psi) relative (external compressed air)
Necessary quality of air: Class 3 from ISO 8573-1:
max. particle size: 5 μm
max. pressure dew point: –20°C (–4°F)
max. oil content: 1 mg/m3
Environmental conditions (EN 61010): Indoor use only
Air humidity: 10 to 90% relative humidity, non-condensing
Operating temperature: +15 to +35 °C (+59 to +95 °F)
Power supply: supplied by master instrument
Mains voltage according to the type plate at the rear of master instrument (protective grounding required)
Dimensions excl. master instrument (L x W x H): 480 mm x 230 mm x 420 mm (18.9 in x 9.1 in x 16.5 in)
Weight excl. master instrument: approx. 9.4 kg (20.7 lbs)