Radiation Type: X and Gamma - from 24.6 keV to 10MeV
Measured quantities: Ambient dose equivalent rate H*(10), Ambient dose equivalent H*(10). Not for pulsed radiation fields
Measuring range, dose rate: 0.05 mkSv/h ÷ 5 Sv/h (up to 100 Sv/h)
Measurement principle: Dead time independent
Integration time: 10 s … 6.55 · 105 s (configurable)
Overexposure: 50 Sv/h over 5 min without decrease in counting rate
Response time: ≤ 8 s (1 μSv/h … 10 mSv/h, peak finder activated)
: ≤ 2 s (10 mSv/h … 5 Sv/h, peak finder activated)
Data interfaces: RS 485, RS 232, option USB
Operating temperature: −30 °C to +60 °C
Power supply: 10 ÷ 30 VDC, 1.5W
Protection: IP67 (including connector and filling valve)
Enclosure: Ø 100 x 400 mm
Weight: 1.5 kg