HORIBA HP-480 pH meter (0 pH ~ 14 pH)
Measuring range: 0 pH to 14 pH: 0.01 pH resolution.
Temperature: 0℃ to 100℃: 1℃ resolution (selectable display).
Transmission output: 4 mA to 20 mA DC: input/output isolated type.
Maximum load resistance: 900 Ω.
Transmission output range: Free.
Repeatability and linearity: ±0.05 pH, ±0.08 mA (for transmission output range of 0 pH to 14 pH).
Contact output: Outputs: 2 points.
Alarm contact output (R1 and R2).
Contact type: Relay contact, SPDT (1c).
Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load).
Contact function: Selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance.
Calibration function: Two-point automatic calibration or manual calibration.
Two-point automatic calibration: With automatic validation of the electrode characteristics.
Types of standard solution: pH 2, pH 4, pH 7, pH 9 and pH 10 (JIS).
Combination of standard solutions: pH 7 and one of the others.
Manual calibration: Freely selectable, but the difference should be over 2 pH.
Temperature calibration: One point.
Transmission output hold feature:
Previous value hold.
Optional value hold.
Continuous.
Selectable from the above (however, only the previous value hold is available in the maintenance mode).
Self-diagnosis function:
Calibration function: Asymmetry potential error, sensitivity error, response speed error and standard solution error.
Electrode self-diagnosis: Temperature sensor short-circuit and temperature sensor disconnection.
Outside of the measuring range.
Converter error.
Temperature compensation element: Selectable from compensation ON (350 Ω (25℃), 500 Ω (25℃), 1 kΩ (0℃), 6.8 kΩ (25℃) and 10 kΩ (25℃)) and compensation OFF.
Temperature-compensated range: 0℃ to 100℃.
Ambient temperature: −5℃ to 45℃.
Relative humidity: 20% to 85% (without dew condensation).
Storage temperature: −25℃ to 65℃.
Power supply: Rated voltage 100 V AC to 240 V AC, 50 Hz/60 Hz, 10 VA (max.).
Construction: Panel-mounting type for indoor use.
Panel case: ABS.
Terminal: PBT.
Panel: Water-resistant structure.
Protection structure:
Front panel: IP65 (IEC60529, JIS C0920).
Rear case: IP20.
Terminal: IP00.
Class II equipment (IEC61010-1).
Pollution degree 2 (IEC61010-1).
Applicable standards:
CE certification.
EMC Directive: EN61326-1.
Low Voltage Directive: EN61010-1.
RoHS Directive: EN50581.
Immunity (Industrial electromagnetic environment).
Noise increase: ±0.2 pH or less.
FCC Rules: FCC Part 15.
External dimensions: 48 (W) × 96 (H) × 115 (D) mm.
Case depth: Approx. 105 mm.
Weight: Approx. 400 g.
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts