Volume: 108 liters
Interrior dimensions: 560W x 480H x 400D mm
Exterior dimensions: 745W x 864H x 584D mm
Stainless steel grids: 2
Max. number of grids: 5
Max. loading per grid: 20 kg
Max. loading of chamber: 175 kg
Working temperature range: min. 5 above ambient temperature to 80°C
Setting accuracy: 0.1°C
Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable temperature values
adjustable parameters: temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime
Sterilisation: fixed sterilisation programme (4 hours/160°C) for sterilisation of working chamber, not for sterilising the load
Ventilation
Convection: natural convection
Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
Vent: vent connection with restrictor flap
Communication
Documentation: programme stored in case of power failure
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Ambient temperature: 5 – 40°C
Humidity: max. 80% rh, non-condensing
Voltage: 230 V, 50/60 Hz
Electrical load: 1400 W
Installation: mounted on lockable castors
Standard equipment
Works calibration certificate: incl. works calibration certificate for +37°C
Door: inner glass doors
Door: fully insulated stainless steel doors with2-point locking(compression door lock)
Internals: 2 stainless steel grid(s), electropolished