Volume: 32 liters
Interrior dimensions: 400W x 320H x 250D mm
Exterior dimensions: 585W x 704H x 434D mm
Stainless steel grids (standard): 1
Max. number of grids: 3
Working temperature range: at least 5 above ambient temperature to 80℃
Setting accuracy: 0.1℃
2 Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking overfunctions in case of an error
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour display.
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function setpointWAIT: the process time dose not start until the set temperature is reached
Calibration: three freely selectable temperature values
Adjustable parameters: temperature (Celsius or Fahrenheit), programme time, time zones
Convection: natural convection
Programme stored in case of power failure
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Temperature control: mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature, overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
Additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system: for fault analysis
Alarm: visual and acoustic
Door: fully insulated stainless steel door with 2-point locking
Structured stainless steel housing, scratch-resistant, robust and durable; rear of zinc-plated steel
Internal data logger with a storage capacity of at least 10 years
Voltage: 230 V, 50/60 Hz
Electrical load: 1600 W
Ambient temperature: 5 – 40℃
Humidity: max. 80% rh, non-condensing