Insulation resistance

Insulation resistance (test voltages 50 V DC, 100 V DC and 250 V DC)

Measuring range according to EN 61557 is 0.25 MΩ ÷ 199.9 MΩ

Measuring range (MΩ)/ Resolution (MΩ)/ Accuracy

0.00 ÷ 19.99/ 0.01/ ±(5 % of reading + 3 digits)

20.0 ÷ 99.9/ 0.1/ ±(10 % of reading)

100.0 ÷ 199.9/ 0.1/ ±(20 % of reading)

Insulation resistance (test voltages 500 V DC and 1000 V DC)

Measuring range according to EN61557 is 0.15 MΩ ÷ 30 GΩ

Measuring range (Ω)/ Resolution (MΩ)/ Accuracy

0.00 M ÷ 19.99 M/ 0.01/ ±(5 % of reading + 3 digits)

20.0 M ÷ 199.9 M/ 0.1/ ±(5 % of reading)

200 M ÷ 299 M/ 1/ ±(5 % of reading)

300 M ÷ 999 M/ 1/ ±(5 % of reading)

1.00 G ÷ 4.99 G/ 10/ ±(10 % of reading)

5.00 G ÷ 19.99 G/ 10/ ±(20 % of reading)

20.0 G ÷ 29.9 G/ 100/ indicative

Open circuit voltage: -0 % / +20 % of nominal voltage

Measuring current: min. 1 mA at RN=UNx1 kΩ/V

Short circuit current: max. 3 mA

Auto discharge after test

Continuity (200 mA)

Measuring range according to EN61557 is 0.16 Ω ÷ 1999 Ω

Measuring range R (Ω)/ Resolution (Ω)/ Accuracy

0.00 ÷ 19.99/ 0.01/ ±(3 % of reading + 3 digits)

20.0 ÷ 199.9/ 0.1/ ±(5 % of reading)

200 ÷ 1999/ 1/ ±(5 % of reading)

Open-circuit voltage: 6.5 V DC ÷ 9 V DC

Measuring current: min. 200 mA into load resistance of 2 Ω

Test lead compensation: up to 5 Ω

Automatic polarity reversal of the test voltage.

Continuity (7 mA)

Measuring range (Ω) Resolution (Ω) Accuracy

0.0 ÷ 19.9 0.1 ±(5 % of reading + 3 digits)

20 ÷ 1999 1 ±(5 % of reading + 3 digits)

Open-circuit voltage 6.5 V DC ÷ 9 V DC

Short-circuit current max. 8.5 mA

Test lead compensation up to 5 Ω

Voltage

Measuring range (V)/ Resolution (V)/ Accuracy

0 ÷ 9.99/ 0.01/ ±(3 % of reading + 3 digits)

10.0 ÷ 99.9/ 0.1/ ±(5 % of reading + 3 digits)

100 ÷ 550/ 1/ ±(5 % of reading + 3 digits)

Result type: True r.m.s. (TRMS)

Nominal frequency range: 0 Hz, 15 Hz ÷ 500 Hz

Frequency

Measuring range (Hz)/ Resolution (Hz)/ Accuracy

0.00 ÷ 19.99/ 0.01/ ±(0.2 % of reading + 1 digit)

20.0 ÷ 199.9/ 0.1/ ±(0.2 % of reading + 1 digit)

200 ÷ 500/ 1/ ±(0.2 % of reading + 1 digit)

Nominal voltage range: 10 V ÷ 550 V

General data

Power supply: 6 x 1.5 V batteries or rechargeable batteries, size AA

Battery life: typical 13 h (>1200 insulation resistance tests / >2000 resistance

RLOW tests)

Overvoltage category: CAT III 600 V, CAT IV 300 V

Protection classification:IP 40

Display: custom liquid crystal display with backlight

Communication transfer speed: 115200 baud (RS 232) / 256000 baud (USB)

Memory: 1900 measurements

Dimensions (w x h x d): 14 cm x 8 cm x 23 cm

Weight (without batteries): 0.80 kg 

MI 3121 Standard set

Instrument Smartec Insulation / Continuity

Soft hand strap

Test lead, 2 x 1.5 m

Test probe, 2 pcs (black, red)

Crocodile clip, 2 pcs (black, red)

Power supply adapter + 6 NiMH rechargeable batteries, type AA

Instruction manual on storage media

Short instruction manual

Instruction manual

Datasheet

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts