Detector measuring force: 4 mN
Stylus tip angle: 90°
Stylus tip radius: 5 µm
Traverse: 25 mm
Measuring speed: 0.05, 0.1, 0.5, 1 mm/s
Measuring method: Skidless / skidded
Range: 800 µm; 80 µm; 8 µm (up to 2.4 mm with an optional stylus)
Positioning: ±1.5° (tilting), 10 mm (up/down)
Profiles: Primary Profile (P), Roughness Profile (R), Waviness (W), MOTIF (R, W) and more
Standard: EN ISO, VDA, JIS, ANSI and customize settings
Analysis graphs: BAC, ADC
Digital filter: Gauss, 2CR75, PC75
No. of sampling length (L): x 1, x 3, x 5, x L* (* = oder jeder beliebige Wert)
Cut-off length:
λc : 0.08 mm; 0.25 mm; 0.8 mm; 2.5 mm; 8 mm
λs : 2.5 µm; 8 µm; 25 µm
Printer: Thermal Printer
Tolerance: Coloured upper / lower limit
Interface: USB, Digimatic, RS-232C, Foot switch
Power supply: Via AC adapter, rechargeable battery
Functions:
Traverse : 25 mm
Traverse straightness : 0.3 µm / 25 mm