Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Size Range: 0.5 μm to 25 μm
Size Channels: Factory calibrated at 0.5, 0.7, 1.0, 2.5, 5.0, 10.0 μm
Counting Efficiency: 50% @ 0.3μm; 100% for particles > 0.45 μm per JIS
Flow Rate: 0.1 CFM (2.83 LPM)
Concentration Limits: 10,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4), 20,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Light Source: Long life laser diode
Zero Count: <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction. No appreciable drift.
Alarms: Channel alarms on Raw counts, concentrations or mass (alarms on optional environmental sensors)
Calibration: NIST traceable
Vacuum Source: Internal vacuum pump with HEPA filter
Filtered Exhaust: Internal HEPA filter
Airflow: Internally monitored
Number of Channels: 6 channels (up to 30 channels of simultaneous data)
Configuration/Download: USB mini-B
Alarm: Alarm LED ring
Communication Modes: MODBUS™ RTU or ASCII outputs (over isolated RS-485), TCP/IP, PoE, or WiFi
Optional Environmental Sensor: Temperature and Relative Humidity probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC), 15-90% ±2% RH
Standards: ISO 21501-4 and JIS B9921
Instrument Calibration: Recommended minimum once per year
External Surface: Stainless steel
Dimensions (L x W x H): 3.59 x 1.80 x 4.86 in (9.1 x 4.5 x 12.3 cm)
Weight: 1.71 lb. (780 grams)
Accessories: Terminal block plug, User Manual, and Instrument Management Software (USB Key)
Optional Accessories: Printed manual, barb fittings, mounting bracket, sample tubing, zero-count filter, and temp./RH probe
Buffer Memory: 65,000 sample records (rotating buffer) including particle count data and environmental data
Sample Time: 1 second to 99 hours
Power: 9 - 24 VDC (< 3.5 watts)
Operating Conditions: 41-104ºF (5-40ºC) / 20% to 95% non-condensing
Storage Conditions: 32-122ºF (0-50ºC) / Up to 98% non-condensing