INPUT
- Voltage: 100 - 240 VAC ± 10% Full Range
- Frequency: 50/60 Hz ± 5%
- Fuse: 12A / 250 VAC Slow-Blow
GROUND BOND TEST MODE
- Output rating: 3.0 - 60.0 AAC
- Voltage setting: 0-5.00 kVAC; 0-6.00 kVDC
+ Resolution: 0.1A
+ Accuracy: ± (2% of setting + 0.1A)
+ Voltage 12 VAC (fixed)
- Output Frequency: 50/60 Hz user selectable
+ Accuracy: ± 0.1%
- Resistance limit settings: 0 - 150 mΩ for 30.1 - 60.0 A
0 - 200 mΩ for 15.1 - 30.0 A
0 - 600 mΩ for 3.0 - 15.0 A
+ Resolution: 1 mΩ
+ Accuracy: ± (2% of setting + 2 mΩ)
- Offset limit settings: 0 - 100 mΩ
+ Resolution: 1 mΩ
+ Accuracy: ± (2% of setting + 2 mΩ)
- Dwell timer: 0, 0.5 - 240.0 sec, (0=continuous), 0.1 sec/step
- Ramp timer: 0.1 sec fixed
- Measurement current: 0.0 - 60.0 AAC
+ Resolution: 0.1 A
+ Accuracy: ± (3% of reading + 0.1 A)
- Ohmmeter: 0 - 600 mΩ
+ Resolution: 1 mΩ
+ Accuracy: ± (3% of reading + 3 mΩ) for 3 - 5.9 A,
± (2% of reading + 2 counts) for 6 - 60A
GENERAL SPECIFICATIONS
- Memories: 5
- Remote I/O:
+ Input: Test, Reset, Interlock
+ Output: Pass, Fail, Test-in-Process
- Voltage limit settings: 0.00 - 9.00 VAC
+ Resolution: 0.01 V
+ Accuracy: ± (2% of setting + 0.02 V)
- Offset limit settings: 0.00 - 6.00 VAC
+ Resolution: 0.01 V
+ Accuracy: ± (2% of setting + 0.02 V)
- Safety mark: CE/cTUVus
- Dimensions (WxHxD): 16.93" x 5.20" x 11.81" (430 x 132 x 300 mm)
- Weight: 20.25 lbs. (9 Kg)