DC voltage
Input type/ Measuring range (indicating range)/ Measurement accuracy (reference error, absolute error)/ Digital display resolution
0-10V/ -0.5V to +11.000V/ 0.01% F.S. ±0.002V/ 0.01-0.001V(optional)
0-5V/ -0.5V to+5.500V/ 0.01% F.S. ±0.002V/ 0.01-0.001V(optional)
±100mV/ -110.0mV to+110.0mV/ 0.001% F.S. ±0.002mV/ 0.01-0.0001V(optional)
±20mV/ -21.0mV to+21.0mV/ 0.001% F.S. ±0.002mV/ 0.01-0.0001V(optional)
DC current
Input type/ Measuring range (indicating range)/ Measurement accuracy (reference error, absolute error)/ Digital display resolution
4-20mA/ +3mA to +21.00mA/ 0.01% F.S. ±0.001mA/ 0.01-0.0001V(optional)
Thermocouple
Input type/ Measuring range (indicating range)/ Measurement accuracy (relative error)/ Digital display resolution
K/ -60°C to +1372°C/ ±(0.05% rdg. +0.5°C)/ 0.01°C
J/ -200°Cto+1200°C/ ±(0.05% rdg. +0.5°C), ≤0°C ±(0.15%rdg +0.5°C)/ 0.01°C
E/ -200°C to +1000°C/ ±(0.05% rdg. +0.5°C), ≤0°C ±(0.15%rdg +0.5°C)/ 0.01°C
T/ -200°C to +400°C/ ±(0.05% rdg. +0.5°C), ≤-30°C ±(0.15%rdg +0.5°C)/ 0.01°C
N/ -200°C to +1300°C/ ±(0.05% rdg +0.7°C), ≤0°C ±(0.3%rdg.+0.7°C)/ 0.01°C
W/ +1500°C to +2315°C, 0°C to +1500°C / ±(0.05% rdg. +1.5°C) , ±(0.05% rdg. +1.0°C)/0.01°C
R/ +800°C to +1768°C, +400°C to +800°C/ ±(0.05% rdg +1.0°C), ±(0.2% rdg +2.0°C)/ 0.01°C
S/ +800°C to +1768°C, +400°C to +800°C/ ±(0.05% rdg. +1.0°C), ±(0.2% rdg. +2.0°C)/ 0.01°C
B/ +800°C to +1820°C, +400°C to +800°C/ ±(0.05% rdg +1.0°C), ±(0.2% rdg. +2.5°C)/ 0.01°C
Thermal resistance
Input type/ Measuring range (indicating range)/ Measurement accuracy (relative error)/ Digital display resolution
Pt100/ -200°C to +660°C/ ±(0.05% rdg. +0.3°C)/ 0.01°C
Cu50/ -50°C to +150°C/ ±(0.05% rdg. +0.3°C)/ 0.01°C
PT1000/ -200°C to +300°C/ ±(0.05% rdg +0.2°C)/ 0.01°C
Preheat time: over 30 minutes
Thermocouple measurement: Includes cold junction compensation accuracy
Standard operating: Temperature 25±3°C, humidity 55±10%RH
Environmental adaptability
Operating temperature: -20°C to +60°C
Working humidity: 0 to 95%RH (no condensation)