emin

GW instek GPT-9904 Safety Testers (5kVAC, 6kVDC, IR, GB, 500VA)

  • Manufacturer: GW INSTEK
    Model: GPT-9904
    • Request a Quotation Request a Quotation
  • Price (exclude tax) 105,515,000 VND
    Price (include tax) 116,066,500 VND
  •  Call for best price

    Hotline: (+84) 867820758

    Phone+84 (24)62923267

AC WITHSTANDING

Output-Voltage Range

0.100kV~ 5.000kV ac

Output-Voltage Resolution

2V/step

Output-Voltage Accuracy

± (1% of setting +5V) [no load]

Maximum Rated Load

500 VA (5kV/100mA)

Maximum Rated Current

100mA (0.5kV < V 5kV)

10mA (0.1kV V 0.5kV)

Output-Voltage Waveform

Sine wave

Output-Voltage Frequency

50 Hz / 60 Hz selectable

Voltage regulation

± (1% of reading +5V) [full load no load]

Voltmeter Accuracy

± (1% of reading+ 5V)

Current Measurement Range

0.001mA~100.0mA

Current Best Resolution

0.001mA / 0.01mA / 0.1mA

AC Current Measurement Accuracy

±(1.5% of reading +30 counts) when HISET < 1.11mA

±(1.5% of reading +3 counts) when HISET 1.11mA

Window Comparator Method

Yes

ARC Detect

Yes

RAMP (Ramp-Up Time)

0.1~999.9s

TIMER (Test Time)*

OFF, 0.5s~999.9s

Sweep Function*

Yes

GND

ON/OFF

DC WITHSTANDING

Output-Voltage Range

0.100kV~ 6.000kV dc

Output-Voltage Resolution

2V/step

Output-Voltage Accuracy

± (1% of setting +5V) [no load]

Maximum Rated Load

100W (5kV/20mA)

Maximum Rated Current

20mA (0.5kV < V 6kV)

2mA (0.1kV V 0.5kV)

Voltage regulation

± (1% of reading +5V) [full load no load]

Voltmeter Accuracy

± (1% of reading+ 5V)

Current Measurement Range

0.001mA~020.0mA

Current Best Resolution

0.001mA / 0.01mA / 0.1mA

DC Current Measurement Accuracy

±(1.5% of reading +30 counts) when HISET < 1.11mA

±(1.5% of reading +3 counts) when HISET 1.11mA

Window Comparator Method

Yes

ARC Detect

Yes

RAMP (Ramp-Up Time)

0.1~999.9s

TIMER (Test Time)*

OFF, 0.5s~999.9s

Sweep Function*

Yes

GND

ON/OFF

INSULATION RESISITANCE

Output-Voltage

50V~1000V dc

Output-Voltage Resolution

50V/step

Output-Voltage Accuracy

± (1% of setting +5V) [no load]

Resistance Measurement Range

0.001GW~ 50.00GW

 

Test voltage

Measurable Range

 

50VV450V

0.001~ 0.050GW

0.051~ 2.000GW

 

500V

Chi tiết

Download Datasheet


  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information