For full functionality of this site it is necessary to enable JavaScript.

Mecmesin OmniTest Touch 50 Universal Testing Machine (50 kN / 11000 lbf, Console-driven)

Load

Rated capacity: 50kN/ 5000kgf/ 11000lbf

Number of ballscrews: 2

Max data acquisition rate: 1000Hz

Enhanced Load Cells (ELS)

Accuracy: When calibrated as part of a system, accuracy is ±0.5% of reading down to 2% of range, suitable for use with Class 0.5 systems according to the requirements of ISO 7500-1. Eg, ELS 100 when measuring 10N is accurate to ± 0.05 N or when measuring 50N it is accurate to ± 0.25 N.

Resolution1: 50,000 filtered from 24 bit

Units of measurement: mN, daN, N, kN, gf, kgf, ozf, lbf

Sampling rate: 20 kHz

Data-acquisition rate: 1000Hz

Displacement

Crosshead travel (maximum between limit switches): 1230 mm

Accuracy: ±0.1% of indicated position or ± 0.01 mm, whichever is greater

Resolution: 0.001mm

Speed

Range: 0.01 - 500 mm/min (0.0004 - 19.7"/min)
Accuracy: Better than ±2% of indicated speed or ± 0.02 mm/min (20 µm/min), whichever is greater* * See help.mecmesin.com for additional info

Resolution: 0.001 mm/min (0.00004"/min)

Dimensions: 

Distance between columns: 425 mm (16.75")

Headroom/vertical daylight (vertical space to fit sample): 1330 mm (52")

Height: 1938 mm (75")

Width: 986 mm (39")

Depth: 649 mm (26")

Weight: 442 kg (972 lb)

Common specifications

Voltage: 230V AC 50 Hz or 110V AC 60 Hz

Power (max): 750W

Operating temperature: 10 - 40°C (50 - 104°F)

Humidity range: Normal industry and laboratory conditions, non-condensing 30 - 80% RH

Details

Brochure

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information