Ống phát hiện khí hoặc hơi Formaldehyde HCHO Gastec 91P (0.02-1.44 ppm)
Hãng sản xuấtGastec(Xem thêm sản phẩm của hãng)
Model91P
Liên hệ
Thanh toán an toàn
Hỗ trợ chuyên nghiệp
Đổi trả dễ dàng
Giao hàng tận nơi
Thang đo hiệu chuẩn: 0.02-0.4 ppm
Thang đo: 0.02-1.44 ppm
Tốc độ lấy mẫu: 200mL/min
Thể tích lấy mẫu: 6000mL, 2000mL
Thời gian lấy mẫu: 30 phút, 10 phút
Chuyển đổi màu: vàng sang hồng
Số ống trong hộp: 10
Tương thích với bơm lấy mẫu tự động: GSP-300FT-2
Specifications
| Calibration Range | 0.02-0.4 ppm |
| Measuring range | 0.02-1.44 ppm |
| Sampling (flow) Rate | 200mL/min |
| Sampling Volume | 6000mL, 2000mL |
| Sampling Time | 30min, 10min. |
| Detecting Limit | 0.01 ppm |
| Colour Change | Yellow → pink |
| Corrections for temperature/humidity | Temperature |
| Shelf life | 1 year (in the refrigerator) |
| Accuracy | CV=5-10% (CV: Coefficient of variation = σ: standard deviation / mean value x 100) |
Possibly coexisting substances and their interferences (NOTE)
| Substance(s) | Interference | Colour change by itself |
| Ammonia | 2 ppm (or less) | No | No |
| Nitrogen dioxide | 1 ppm (or less) | No | No |
| Acetaldehyde | "+" error (higher reading) |
produces pinkish discolouration | |
| Acetone | 0.3 ppm (or less) | No | Produces pinkish stain around tube opening (inlet) |
| Ethanol | No | No | |
| Ethyl acetate | No | No | |
| p-dichlorobenzene | No | No | |
| Toluene | No | No | |
Scrubber 1 (black): removes acetone Scrubber 2 (purple): Ammonia and nitrogen dioxide are removed. The presence of ammonia causes the purple scrubber 2 to turn yellow. The presence of nitrogen dioxide causes scrubber 2 to turn a darker shade of purple |
|||
| Calibration gas generation | Diffusion tube method (analysis: by absorptiometry, acetylacetone method) |
Miscellaneous
| Electric sampling pump This detector tube is suited for measuring formaldehyde in indoor environments. The GSP-200 / GSP-300FT series automatic gas samplers (or comparable sampling devices/pumps) should always be used for optimal performance. A deviation of +/-10% in the sampling flowrate (i.e. amount of air flowing through the tube at a given moment) or the sampling amount (total sampled amount of target substance) causes an average margin of error of approx. +/-10% (maximum +/-20%) in the indicated value. Formaldehyde indoor concentration threshold: 0.08 ppm (mean value during 30 minutes) |
| TLV-TWA | TLV-STEL | Explosive range |
| -- -- | C0.3ppm | 7.0。チ73% |