For full functionality of this site it is necessary to enable JavaScript.
EMIN.VN
0
Product image

Gastec 7J ホスフィンクイック測定検出器チューブ PH 3(2.5〜1000 ppm)

お問い合わせ
安全なチェックアウト
高品質な対応
簡単な交換・返品
配送対応可能

Indicator: 50 to 500 ppm
Measuring Range: 
+ 2.5 to 25 ppm, Number of Pump Strokes 2 to 10
+ 25 to 500 ppm , Number of Pump Strokes 1
+ 500 to 1000 ppm, Number of Pump Strokes 1/2
Sampling Time:  1 minutes per pump stroke
Colour Change: White → Pale yellow
Reaction Principle: 2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl23H2O+6HCl
Coefficient of Variation: 10% (for 2.5 to 100 ppm), 5% (for 100 to 500 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Performance
Measuring Range
2.5 to 25 ppm 25 to 500 ppm 500 to 1000 ppm
Number of Pump Strokes
2 to 10 1 1/2
Correction Factor
1/2 to 1/10 1 2
Sampling Time 1 minute per pump stroke
Detecting Limit 1 ppm (n = 10)
Colour Change White → Pale yellow
Reaction Principle 2PH3 + 6HgCl2 + 3H2O → Hg3P2 3HgCl23H2O+6HCl
Coefficient of Variation 10% (for 2.5 to 100 ppm), 5% (for 100 to 500 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Arsine 1/3 time + Brown
Hydrogen selenide 1/5 time + Brown
Hydrogen sulphide 1/5 time + Dark brown
Calibration gas generation High pressure gas cylinder method
TLV-TWA TLV-STEL Explosive range
0.3 ppm (2014) 1 ppm (2014) -

 

お得な情報を受け取る

数量割引、まとめ買い価格の更新、新製品情報をメールでお届けします。

登録することで、当社の利用規約およびプライバシーポリシーに同意したものとみなされます。

クイックサポート

認定専門家へ直接アクセス