ACW
+ Out voltage range: 500~3000VAC,10V/step, ±(2.5%×setting+10V)
+ Output frequency: 50/60Hz, ±0.1%×Setting
+ Breakdown current range: 0.100~20.00 mA, ±(2.5%×reading+5 counts)
IR
+ Output voltage range: 500~3000VAC,100V/step; ±5%×setting
+ Energy capatitor:Standard 0.255uH
+ Sampling frequency:50kHz~40MHz,15 adjustemnt range
+ Test time:1~10 times continuous
IW
+ Pulse voltage range: 500~3000VAC,100V/step; ±5%×setting
+ Energy capatitor:Standard 0.022uH
+ Sampling frequency: 50kHz~40MHz,15 adjustment range
Power/Voltage start
+ Voltage range: 20~460V,10V/step; ±(0.3%×reading +0.2% ×range)
+ Current range: 0.05~6.00~40.00A; ±(0.3%×reading + 0.2% range)
+ Active pwer range: 1.0~99.9W~18kW; ±(0.8%×reading +0.2×range)
+ Frequency range:45~65Hz; ±0.1%×reading
Weight: 33 kg
Product Detail
Main features
* Lock rotor without load:Adopt low voltage start to test lock current,power,power input adjustement automatic,sample operation,precision test
* Sensitive roation:Adopt contactless test method,sample operation
* Correction:3 phase current and power correction,3 phase resistance and current balance calculation
* 1/3 phase: single and 3 hpase test mode
* Large ACW capactiy: up to 5KW,100Ma
* Large energy IW:Upto 0.1uF capactitor
* Multi-high voltage output:Case high voltage output,return wires high voltage output,P-P output are optional
Model |
AN96951(F) |
AN96950(F) |
AN9653(F) |
ACW |
Out voltage range:500~3000VAC,10V/step.(5000V can be modified) Voltage accuracy:±(2.5%×setting+10V) Output frequency:50/60Hz Frequency accuracy:±0.1%×setting Breakdown current range:0.100~20.00 mA(100Ma can be modified) Current accuracy:±(2.5%×reading+5 counts) |
||
IR |
Output voltage range:500~3000VAC,100V/step.(5000V can be modified) Voltage peak accuracy:±5%×setting Energy capatitor:Standard 0.255uH,0.047uH is optinal Front time of impulse:0.5uS Sampling frequency:50kHz~40MHz,15 adjustemnt range Test time:1~10 times continuous Ajustment:Area,differ |
||
IW |
Pulse voltage range:500~3000VAC,100V/step.(5000V can be modified) Voltage peak accuracy:±5%×setting Energy capatitor:Standard 0.022uH,0.047uH is optinal Front time of impulse:0.5us Sampling frequency:50kHz~40MHz,15 adjustment range Test time:1~10 times continuous Ajustemnts:Area,differ |
||
DCR |
---- |
Range:0.020~0.200~2.000~20.00~200.0~2000~20000Ω Accuracy:±(0.3%×reading + 3 counts) |
--- |
Power/Voltage start |
--- |
Voltage range: 20~300V,10V/step. Accuracy:±(0.3%×reading +0.2%×range) Current range: 0.20~2.00~20.00A Accuracy:±(0.3%×reading + 0.2%×range) Active pwer range: 1.0~99.9W~6KW Accuracy:±(0.8%×reading +0.2%×range) Frequency range:45~500Hz Accuracy:± 0.1% ×reading |
Voltage range: 20~460V,10V/step. Accuracy:±(0.3%×reading +0.2% ×range) Current range: 0.05~6.00~40.00A (100Ma can be modified) Accuracy:±(0.3%×reading + 0.2% range) Active pwer range: 1.0~99.9W~18kW Accuracy:±(0.8%×reading +0.2×range) Frequency range:45~65Hz Accuracy:±0.1%×reading |
C voltage |
--- |
Voltage range:60.0~600.0V Accuracy:±0.6%×reading |
--- |
Speed |
--- |
Range:120~99,999rpm Accuracy:±0.1%×reading+1 |
--- |
Rotation |
--- |
RW/RS/RI/KEEP |
--- |
PG single |
--- |
PG power:5V or 12V Voltage test:0.50~15.0V Accuracy:±(0.5×reading +2 counts) Frequece range:10~1000Hz Accuracy:±0.1%×reading+1 count |
--- |
Weight(without packaging) |
33Kg |