For full functionality of this site it is necessary to enable JavaScript.

ESCO CCL-240T-9-HHS CelCulture® CO₂インキュベーター (240 L, 180°C HHS, 115 VAC 50/60 Hz)

SUPPRESSED O2 MODEL WITH STAINLESS STEEL CHAMBER

CelCulture® Incubator 240L IR Sensor, CO2/O2 Control, ULPA, 180°C HHS, 115 VAC 50/60 Hz


O2: 

O2 Control System: Microcontroller PI

O2 Control Range, % O2: 1-18

O2 Fluctuation, % O2: ± 0.3

O2 Sensor: Zirconia O2 Sensor


TEMPERATURE: 

Temperature Control Method: Direct Heat and Air Jacket using Microcontroller PI

Ambient Temperature Range: 18 to 32°C (64.4 to 89.6°F)

Temperature Control Range, °C: Ambient +5 to 60

Temperature Uniformity, °C: Standard Unit: <± 0.35; Suppressed O2 model: <± 0.4

Temperature Fluctuation, °C: ±0.2

Temperature Recovery Time (after 30 seconds door opening, 98% from initial value): ≤5 minutes


CO2: 

CO2 Control System: Microcontroller PI

CO2 Control Range, % CO2: 0-19.5

CO2 Fluctuation, % CO2: ± 0.3

CO2 Sensor: Infrared (IR) Sensor

CO2 Recovery Time (after 30 seconds door opening, 98% from initial value): At 5.0% CO2 by volume (Standard unit): ≤5 minutes; Suppressed O2 model: ≤8 mins.


HUMIDITY:

Humidification Method: Humidity pan

Humidity Range, %RH (at 37°C): 85-90


PHYSICAL CONSTRUCTION: 

Interior Volume: 240 L

External Dimensions (W x D x H): 750 x 770 x 906 mm

Internal Dimensions (W x D x H): 595 x 640 x 633 mm

Net Weight: 121 kg

Chamber Construction: Main Body Electrogalvanized steel with Isocide™ antimicrobial coating; Interior Material Stainless steel, type 304

Number of Shelves: 4

Maximum Number of Shelves: 7

Electrical Configuration 110-130 VAC, 50/60 Hz

Full Load Amps: 14 A 

Electrical Configuration 220-240 VAC, 50/60 Hz

Full Load Amps: 7 A


CONTAMINATION CONTROL

Contamination Control Methods

1) Main body is electrogalvanized steel with Isocide™ antimicrobial coating;

2) 180˚C high heat sterilization cycle;

3) ULPA filter (optional) - filter must be removed during sterilization cycle

4) 0.2 µm gas inlet filter

5) 1-micron air circulation filter


All data recorded were observed with unloaded chambers and under optimum factory setting of 22 ±3°C with room humidity of 30-60%.

*Results are achieved when tested at 37°C as set point. Results may vary if set point changes and calibration is needed.

**For temperature not exceeding 37.2°C.

***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.

****For CO2 level not exceeding 5.2%.

*****Results are achieved when tested at 5% O2 as set point. Results may vary if set point changes and calibration is needed.

****** For O2 level not lower than 4.8%.

******* Esco does not guarantee condensation-free chamber at humidity level higher than 90%