For full functionality of this site it is necessary to enable JavaScript.

JEIOTECH TH-ICH-800 温度及び湿度チャンバー (760L,230V/50Hz)

  • 生産者:JEIOTECH
    Model: TH-ICH-800
    P/N: AAHC1012K
    • 要件クオート Request a Quotation
  • 連絡先
    • カートへ追加
    • 要件クオート
  • 校正検証修理サービスを参照してください

      いい価格のため、連絡しなさい

    ホットライン: (+84) 968060220

    電話番号: +84 (24) 62923267

Chamber Volume (L / cu ft)    760 lit

Controller      PID control (LCD Type)

Temperature: -5 to 85 / 23 to 185 without Light

            Fluctuation 1) (±℃ / oF) at 40℃ / 60%RH    0.3 / 0.55

            Variation 1) (±℃ / oF) at 40℃ / 60%RH    0.5 / 0.9

Humidity: 20 to 85 %

             :Fluctuation (±%RH) at 60%RH / 40℃    3

             : Variation 2) (±%RH) at 60%RH / 40℃    5

Light Source (ICH Q1B option 2): Cool White, Ultraviolet-A, Light Uniformity

Time: 25min (20℃ → 85℃)/ 35min (20℃ → -5℃)

Refrigeration: System Air - cooled, R-404A, Capacity (W)   470

Material: Ni-Cr Wire Heater / SUS Tube Heater , Polyurethane 

Interface Port    Basically RS-232C, [RS-485 : optional]

Dimension (W×D×H): Internal (mm / inch)    750×880×1160mm

                                : External (mm / inch)   1220×1215×1965mm

Net Weight (kg / lbs)    400±10 

Electrical Requirements AC230V, 1ph, 50Hz    23.5 A  Cat.No.    AAHC1012K

             

Details

Temperature & Humidity Control range

TG, ICH Graph
TH-TG (190 to 960L)
  • 40~90% RH at 10℃
  • 20~95% RH at 30℃
  • 10~95% RH at 50~80℃
  • 10~80% RH at 90℃
TH-TG (1500L)
  • 40~90% RH at 15℃
  • 30~95% RH at 30℃
  • 20~95% RH at 50~80℃
  • 20~85% RH at 85℃
TH-ICH (300, 760L)
  • 40~80% RH at 15℃
  • 30~85% RH at 30℃
  • 20~85% RH at 50~85℃

Constructional Features

  • High quality #304 stainless steel exterior and interior.
  • Casters for easy mobility when installing or moving environmental test chamber.
  • Convenient maintenance
    - Available to check total operating time.
  • Unique door lock & trim.
  • Vice lock type door and foam silicone packing to ensure a perfect seal.
  • Cable port for external probes or wires.
  • 50mm diameter cable port comes standard on the left side of the unit. Additional 50mm or 80mm diameter cable ports can be custom ordered. (optional)
  • Container type tank for easy adding and changing of water.
    - Maintenance of the water tank can be easily performed from the front side.
  • Door lock.
  • Cyclic defrost functions does not affect the temperature and humidity control for extended operation. 
    (Under 15℃ set value)
  • Powerful two Sirocco fans maintains a constant airflow in the chamber.

Safety Features

  • Leakage breaker for power supply.
  • Over current protection.
  • Overheat / overcool protection.
  • Switch-off after alarm for over heating.
  • Water empty alarm.
  • Door open alarm.
  • Vice lock door type locking mechanism.

Operating Features

  • Microprocess PID control / Auto-tuning / Calibration.
  • Digital LCD display with touch-sensitive 7 keypads.
  • Easy operation with wide LCD screen.
    - Temperature and humidity / timer settings and operation condition.
  • Program patterns are memorized for 10 years due to a battery built into the PCB.
  • Expansion parameter inputs. (conveniently display by letters)
  • Programming operation.
  • Computer interface.
    - Software provided.
    - Save data in excel format.
    - Storage of program information, backup of value settings / recovery, and temperature / humidity values.
    - RS-232C port as standard offer.
  • Chart recorder. (optional)
    - 6-point dots type recorder with digital display.
    - Digital recorder. (6-channels)
  • In the event of a validation the apparatus correctly controls each light and then transfers accurate data and is equipped with light sensors. (UV light sensor / VIS light sensor)
  • Samples that have to be exposed to the light in which it is completely near the UV energy of less than 200W-hr/m2 is provided to compare directly the results from overall intensity of less than 1.20 million lux-hr, drug substances and reagent.
  • Total fluorescent lamp and UV lamp indicator and operating time indicator.

Safety Features

  • Leakage breaker for power supply.
  • Over current protection.
  • Overheat protection.
  • Switch-off after alarm for over heating.
  • Door open alarm.
  • Water empty.
  • Vice lock door type locking mechanism. (optional)
  • Alarm for completed Photostability Test.
Model TH-ICH-300 TH-ICH-800
Chamber Volume (L / cu ft) 300 / 10.6 760 / 26.8
Controller PID control (LCD Type)
Temperature 1) Range without Humidity (℃ /oF) -5 to 85 / 23 to 185 without Light
0 to 85 / 32 to 185 with Light
with Humidity (℃ / oF) +15 to 85 / 59 to185 with Light
Fluctuation 1) (±℃ / oF) at 40℃ / 60%RH 0.3 / 0.55 0.3 / 0.55
Variation 1) (±℃ / oF) at 40℃ / 60%RH 0.5 / 0.9 0.7 / 1.3
Humidity Range (%RH) FULL RANGE : 20 to 85
40 to 80 at 15℃ (59oF)
30 to 85 at 30℃ (86oF)
20 to 85 at 50℃ to 85℃ (122oF to 185oF)
Fluctuation 2) (±%RH) at 60%RH / 40℃ 3
Variation 2) (±%RH) at 60%RH / 40℃ 5
Light Source
(ICH Q1B option 2)
Cool White 3) 6,000 lux 6,500 lux
Ultraviolet-A 3) 4.5 W/㎡ 5 W/㎡
Light Uniformity VIS. : ±10% UVA : ±15% VIS. : ±15% UVA : ±15%
Typical Time Required to Reach ICH Recommendation of 1.2 million lux-hr ≒ 200Hr ≒ 184Hr
Typical Time Required to Reach ICH Recommendation of 200 W-hr/m2 ≒ 50Hr ≒ 40Hr
Time Heating 4) 25min (20℃ → 85℃) 30min (20℃ → 85℃)
Cooling 4) 35min (20℃ → -5℃) 40min (20℃ → -5℃)
Refrigeration System Air - cooled
Refrigerant R- 404A
Capacity (W) 250 470
Cooler Copper plate fin cooler
Heater Material Ni-Cr Wire Heater / SUS Tube Heater
Drying Bulb (W) 1500 2300
Humidity Bulb (W) 1200 2000
Sensor Temperature Pt 100
Humidity Electronic sensor (Free of drying a wet pack and good response time)
Air Flow 2 set of sirocco fan (2×60 W) : Vertical type
Material Internal SUS #304 (2B polish)
External SUS #304 (Hairline finish) & Steel, 2t, Double painted & baked
Insulation Polyurethane (70mm / 2.76")
Door EPDM Sheet
Operating Temperature 18℃ to 30℃ / 64.4oF to 56oF, 85%RH (Non-corrosive and pure place)
Water Tank Capacity (L / cu ft) 15 / 0.53
Water Quality pH6.2 ~ 7.2, Electrical conductivity 20㎲ / cm to below
Interface Port Basically RS-232C, [RS-485 : optional]
Cable Port Basically one hole provided / Ø 50mm (2"),Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2")
Dimension
(W×D×H)
Internal (mm / inch) 750×650×650 / 29.5×25.6×25.6 750×880×1160 / 29.5×34.6×45.7
External (mm / inch) 1130×985×1520 / 44.5×38.8×59.8 1220×1215×1965 / 48×47.8×77.4
Net Weight (kg / lbs) 280±10 / 617±22 400±10 / 882±22
Electrical Requirements AC230V, 1ph, 60Hz 15.4 A 24.5 A
Cat.No. AAHC1001K AAHC1011K
AC230V, 1ph, 50Hz 14.7 A 23.5 A
Cat.No. AAHC1002K AAHC1012K
1) All specifications are under ambient temperature 20℃ (68oF), No load.
2) Technical data according to DIN 12880, IEC 60068.
3) The value is measured on the center of chamber at the point of vertical direction.
4) Up to 98% of the set value.

 

  • 良質な取り決め
  • オリジナル保証
  • 宅配便
  • 買い取り簡単化