빠른 지원
인증된 전문가에게 직접 연결
인증된 전문가에게 직접 연결
Detector: Scintillation tissue-equivalent plastic, Ø50x40 mm
Energy range: 25 keV – 10 MeV
Measurement range of ambient dose equivalent rate: 30 nSv/h – 1 Sv/h
Measurement range of air kerma rate: -
Limit of intrinsic relative measurement error: ±20%
Energy dependence relative to 662 keV ( 137Cs):
±25% (15 keV – 3 MeV)
±40% (3 – 10 MeV)
Typical sensitivity to gamma radiation, cps/(µSv·h^-1 )
241Am: 3200
60Co: 270
137Cs: 530
Response time for dose rate change from 0.1 to 1 µSv/h (µGy/h): ≤2 s[accuracy error ≤±10%]
Protection class: IР64
Interface: RS232
Power supply: RS232 – 5-9 VDC power source / RS485 – 9-30 VDC power source
Overall dimensions, weight: Ø60х205 mm, 0.5 kg