Division value (d) : 1μg
Maximum range (max) : 2.1g
Repeatability: (5%loading)
Standard deviation: 1.5μg
Typical error: ±3.0μg
Repeatability (close to 100% load)
Standard deviation: 2.5μg
Typical error: ±5.0μg
Linear error standard deviation: 1.5μg
Typical error: ±3.0μg
Position error:
Test weight: 2g
Standard deviation: 1.5μg
Typical error: ±3.0μg
Sensitivity drift (+10℃~+30℃): 5ppm
Verification index value (e) (according to 2014/31/eu specification) : 10μg
Typical stability time: 8s
Typical measurement time: 10s
External calibration weight: 2g
Pan size: ϕ30mm