Controller
It contains the devices for mixing and adjust the combination of propane gas and air
Needle valve to regulate the gas flow
Electronic massive Flow meters type CMOSens
Air: range 0 – 20 NL/min for Nominal Flow 10 l/min at 23°C 0.1 MPa
Propane: range 0 – 3 NL/min for Nominal Flow 650 ml/min at 23°C 0.1 MPa
Accuracy: 0.05% FS
Precision: 0.3% FS (including offset, Nonlinearity, Hysteresis)
Touch LCD display backlit
Gas and Air Pressure gauges dia. 80 mm precision 1%
Automatic measure the time and the rising temperature of the copper block with 100 °C up to +700 °C
Accuracy: time 0.1 seconds
Temperature ± 0.5 °C
Safety Non-return
Burner
According to the Rule Specification IEC 60695-11-2 annex A, with two-position pedestal (V-45°)
Burner 1 kW
Height-adjustable support
For positioning the copper block at 95 mm over the burner
Thermocouple
0.5 mm diameter wire with copper block 9 mm diam, mass ± 0.05 g for the calibration of the flame
Supply requirements
Gas propane purity > 98%
Filtered air: dry and oil-free
230 V – 50 Hz – 30 W
Reference standards
IEC 60695-11-2
UNI EN 60695-11-2