Direct X-ray ruler for X-ray/light field alignment and beam centering in R/F and mammography applications

General Specifications

Range: 5 cm (2 in)

Segment sensors/read out: 41 solid state sensors and corresponding LCD display segments

Segment resolution: 0.25 cm (0.1 in)

Function test: All segments should turn on when the RaySafe DXR+ is fully irradiated.

Power on: Automatic when exposed

Power off: Automatic after 1 min of inactivity

Reset: Automatic

Battery life time: 6-8 years (CR1632) based on 2000 exposures per year

Operating temperature: 10 to 40° C (50 to 104° F)

Storage temperature: 20 to +60° C (-4 to +140° F)

Size (H x W x L): 15 x 30 x 145 mm (0.59 x 1.18 x 5.71 in)

Weight: 75 gr (2.6 oz)

All specifications may change without notice.

Swedish patent 526928.

American patent 7313223.

Recommended Generator Settings

MammoRadiographic
kVpMax5070100
mA>100>200>100>50
FDD (cm)<65<100<100<100

 

Exposure time:
>10 ms No added tube filtration

RaySafe DXR+ Product Leaflet


RaySafe DXR+ User Manual


X-ray Test Equipment Product Catalog